beautypg.com

Brooks Instrument QMBM User Manual

Page 43

background image

Section 4 Maintenance

4-3

Installation and Operation Manual

X-CM-QmB-eng
Part Number: 541B029AAG
June, 2011

QmB Series IP40, IP66, IP66XP

Observation

Cause

Resolution

Oscillating mass flow output

Bubbles dissolved in the liquid.

Use a degasser or use Helium

can cause diagnostic alarm

High pressure push gas can cause

as a push gas with 25-50ft.

bubbles to disolve in fluid. Bubbles

of Teflon

®

tube. Helium is a light

then form at the valve due to

gas and can permeate the

pressure drop, upsetting valve.

Teflon tube to release bubbles.

Oscillating pump pressure used to

Avoid using push gas. Use a pump

move the fluid. Piston action can

or bladder tank to prevent head

cause pressure oscillations,

pressure.

upsetting valve.

Pressure regulators on inlet and
outlet of Quantim to stabilize
pressure variations.

Pressure drop or inlet pressure

Adjust pressure to original

deviates form calibrated values.

specifications.

Valve out of adjustment.

Contact Brooks Instrument.

Unstable inlet pressure.

Check external pressure regulator.

Defective PC board.

Contact Brooks Instrument.

Valves are tuned digitally and

Operate Quantim at specified

mechanically for a particular

ΔP.

pressure conditions.

Valve will not open or fully close

Valves are tuned digitally and

Operate Quantim at specified

mechanically for a particular

ΔP.

pressure conditions.

Proper valve response requires
just the right amount of force to be

Check valve operation with the

applied by the valve. With too

Brooks Service Tool.

much or too little

ΔP the valve

will be mis-tuned.

Mounting attitude is different from

Mount Quantim in the specified

calibration attitude; Reason, gravity

mounting attitude.

effects could pull valve open or
closed.

No Flow or unable to

Insufficient inlet pressure or

Adjust pressures, inspect in-line

achieve full flow

pressure drop.

filters and clean/replace as
necessary

Clogged Valve Orifice or Sensor.

Back flush the device with a
solvent or purge gas with valve
override "Open" (if controller).

Table 4-1 Troubleshooting

This manual is related to the following products: