beautypg.com

6 set modulation and sweep parameters – B&K Precision 4084-4087 - Manual User Manual

Page 24

background image

4080 series Function & Arbitrary/Function Generators –Instruction Manual

24

MODE > START F > STOP F > TIME >TRIG

4.6 Set Modulation and Sweep Parameters

4.6.1. Sweep mode

In the frequency sweep mode, the function generator “steps” from the start frequency to the stop
frequency at a sweep rate which you specify. You can sweep up or down in frequency, and with

either linear or logarithmic spacing. You can also configure the function generator to output a single

sweep (one pass from start frequency to stop frequency) by applying an external trigger. The

function generator can produce a frequency sweep for sine or square waveforms.

Press 【Sweep】 to enable the frequency sweep mode. The display shows a preset frequency (start
frequency) and the “Sweep” symbol appears in the state display area. The carrier frequency will be
displayed and the frequency, amplitude, waveform and DC offset of the carrier signal can be set as

described in previous sections of this chapter. The main functions sine or square wave can be

selected as a carrier signal.

Press the 【menu】key repeatedly to cycle through the sweep menu parameters listed below.



MODE Select LINEAR or LOGarithmic sweep mode

START F sweep start frequency

STOP F: sweep stop frequency

TIME:

sweep time

TRIG:

Select trigger source INTernal or EXTernal

Each time you press the menu key, the parameter will flash for 1 second, followed by the value

of that parameter. Sweep mode parameters can be entered via numerical keypad or the knob.
Once the parameter is set, press 【menu】 to advance to the next parameter.


Sweep mode

MODE parameter: Select LINear (No.1) or LOGarthimic (No.2). In linear sweep mode, the output

frequency changes in a linear fashion during a sweep, whereas in LOG mode the frequency changes

exponentially. The spacing is calculated automatically based on start frequencies, stop frequencies

and sweep time.

Start frequency “START F”:

The frequency where the sweep starts is the start frequency.
After displaying “START F” for 1 second, the current start frequency is displayed automatically and
can be modified via knob or direct data entry. By default, the start frequency will carry over from

the setting for the main wave form
.
Stop frequency “STOP F”
The sweep stops at the frequency “STOP F”
When the start frequency is lower than the stop frequency, the frequency sweep increases gradually

from start frequency (low frequency) to stop frequency (high frequency); When the start frequency

is higher than stop frequency, the frequency sweep decreases gradually from the start frequency

(high frequency) to stop frequency (low frequency).