4080 series Function & Arbitrary/Function Generators –Instruction Manual

22

In case of standard waveforms “sine”, “square”, “triangle”, “rising ramp” and “pulse”, numbers can
be entered and displayed as Peak-to-peak value (Vpp or mVpp) , root mean square value (Vrms and

mVrms) or dBm value.

All other waveforms can only be edited or displayed using Vpp or mVpp units only.

4.5.3 Set DC Offset Voltage

Press【shift】

【offset】to display the current DC offset value. If the current DC offset value is not

equal to zero, the annunciator “Offset” will turn on. The DC offset value can be entered directly or

using the knob.

Example: Set an offset value -1.6V peak-to-peak

Press【shift】

【offset】【-】【1】【●】【6】【Vpp】

or

【shift】【offset】【1】【●】【6】【-】【Vpp】

or

use the knob for value entry

Zero Point Adjustment:

For zero point adjustment of the output signal, using the knob is more convenient than direct entry

via the numerical keypad. The transition of the DC offset Voltage from plus to minus sign will be

automatic when passing through the zero point. The input range of the signal amplitude and DC

offset should satisfy the following equation: |Voffset| + Vpp/2 ≤ Vmax, with the parameters

defined as followed:

Vpp is the peak-to-peak value of the amplitude

|Voffset| is the absolute value of the DC offset

Vmax is 10V at high impedance and 5V at 50Ω load.

The following table shows the corresponding relationship between the Vp-p value of the amplitude

and the absolute DC offset value at high impedance:

Vp-p value of AC signal

Absolute value of DC offset

1.001 V ~ 20.00 V

0 ~ (10.000-Vpp/2) V

316.1 mV ~ 1.000 V

0 ~ 2.000 V

100.1 mV ~ 316.0 mV

0 ~ 632.9 mV

31.01 mV ~ 100.0 mV

0 ~ 200.9 mV

2.000 mV ~ 31.00 mV

0 ~ 62.99 mV


4.5.4 Adjustment of duty cycle

If the current waveform is pulse, and the current display value is amplitude, use the 【Ampl/Duty】

button to toggle between Amplitude and Duty cycle display. When the pulse width is displayed,

enter a value via the numeric keypad or the knob. The valid range is 0.1% ~ 99.9% for frequencies

below 10kHz with a maximum resolution of 0.1%. For frequencies between 10kHz~100kHz the

range is 1% ~ 99% and the maximum resolution is 1%.

Example: Enter a duty cycle value of 60.5%

Press 【Pulse】

【6】【0】【●】【5】【N】

or

use the knob and arrow keys