Atec Rohde-Schwarz-ESRP Series User Manual
Page 9
Version 01.01, January 2013
Rohde & Schwarz R&S
®
ESRP EMI Test Receiver
9
Noise indication (receiver mode),
nominal, calculated from DANL data
RF attenuation = 0 dB, preamplifier = off, termination = 50 Ω, average detector (AV)
R&S
®
ESRP3, R&S
®
ESRP7
9 kHz ≤ f < 100 kHz, BW = 200 Hz
< 0 dB
µV
100 kHz ≤ f < 150 kHz, BW = 200 Hz
< –15 dBµV
150 kHz ≤ f < 1 MHz, BW = 9 kHz
< 2 dBµV
1 MHz ≤ f < 30 MHz, BW = 9 kHz
< –5 dBµV
30 MHz ≤ f < 1 GHz, BW = 120 kHz
< 6 dBµV
1 GHz ≤ f < 3.6 GHz, BW = 1 MHz
< 17 dBµV
R&S
®
ESRP7
3.6 GHz ≤ f < 6 GHz, BW = 1 MHz
< 19 dBµV
6 GHz ≤ f ≤ 7 GHz, BW = 1 MHz
< 21 dBµV
with R&S
®
ESRP-B29 option,
RF attenuation = 0 dB, preamplifier = off, termination = 50 Ω, average detector (AV)
R&S
®
ESRP3, R&S
®
ESRP7
10 Hz, BW = 10 Hz
< 27 dBµV
20 Hz, BW = 10 Hz
< 17 dBµV
100 Hz, BW = 10 Hz
< 7 dBµV
1 kHz, BW = 100 Hz
< 7 dBµV
with R&S
®
ESRP-B2 option,
RF attenuation = 0 dB, preamplifier = on, termination = 50 Ω, average detector (AV)
R&S
®
ESRP3, R&S
®
ESRP7
9 kHz ≤ f < 100 kHz, BW = 200 Hz
< –20 dBµV
100 kHz ≤ f < 150 kHz, BW = 200 Hz
< –25 dBµV
150 kHz ≤ f < 1 MHz, BW = 9 kHz
< –8 dBµV
1 MHz ≤ f < 30 MHz, BW = 9 kHz
< –18 dBµV
30 MHz ≤ f < 1 GHz, BW = 120 kHz
< –7 dBµV
1 GHz ≤ f < 3.6 GHz, BW = 1 MHz
< 5 dBµV
R&S
®
ESRP7
3.6 GHz ≤ f < 6 GHz, BW = 1 MHz
< 7 dBµV
6 GHz ≤ f ≤ 7 GHz, BW = 1 MHz
< 9 dBµV
with R&S
®
ESRP-B2 option and R&S
®
ESRP-B29 option,
RF attenuation = 0 dB, preamplifier = on, termination = 50 Ω, average detector (AV)
R&S
®
ESRP3, R&S
®
ESRP7
1 kHz, BW = 100 Hz
< –13 dBµV
with R&S
®
FSV-B22 option (instruments without R&S
®
ESRP-B2 option only),
RF attenuation = 0 dB, preamplifier = on, termination = 50 Ω, average detector (AV)
100 kHz ≤ f < 150 kHz, BW = 200 Hz
< –20 dBµV
150 kHz ≤ f < 1 MHz, BW = 9 kHz
< –3 dBµV
1 MHz ≤ f < 30 MHz, BW = 9 kHz
< –15 dBµV
30 MHz ≤ f < 1 GHz, BW = 120 kHz
< –4 dBµV
1 GHz ≤ f < 3.6 GHz, BW = 1 MHz
< 7 dBµV
R&S
®
ESRP7
3.6 GHz ≤ f < 6 GHz, BW = 1 MHz
< 9 dBµV
6 GHz ≤ f ≤ 7 GHz, BW = 1 MHz
<11 dBµV
Spurious responses
Image response
30 MHz ≤ f ≤ 7 GHz
f
in
– 2 × 8409.9 MHz (1st IF)
< –80 dBc (nom.)
f
in
– 2 × 729.9 MHz (2nd IF)
< –80 dBc
f
in
– 2 × 89.9 MHz (3rd IF)
< –80 dBc
Intermediate frequency response
30 MHz ≤ f ≤ 7 GHz
1st IF (8409.9 MHz)
< –70 dBc (nom.)
2nd IF (729.9 MHz)
< –80 dBc
3rd IF (89.9 MHz)
< –80 dBc
Residual spurious response
RF attenuation = 0 dB
f ≤ 1 MHz
< –90 dBm
f > 1 MHz
< –103 dBm
Local oscillator related spurious
30 MHz ≤ f ≤ 7 GHz
1 kHz ≤ offset from carrier ≤ 10 MHz
< –70 dBc
offset from carrier > 10 MHz
< –80 dBc
Other interfering signals
Subharmonic of 1st LO
20 MHz ≤ f < 7 GHz,
spurious at 8410 MHz – 2 × f
in
< –70 dBc
Harmonic of 1st LO
mixer level < –25 dBm,
spurious at f
in
–4205 MHz