Preselection and preamplifier (r&s, Esrp-b2 option), Rf preamplifier (r&s – Atec Rohde-Schwarz-ESRP Series User Manual
Page 6: Fsv-b22 option), If and resolution bandwidths
Version 01.01, January 2013
6
Rohde & Schwarz R&S
®
ESRP EMI Test Receiver
Preselection and preamplifier (R&S
®
ESRP-B2 option)
Preselection
State
receiver mode
always on
analyzer mode
on/off (selectable)
Number of preselection filters
16
Bandwidths (–6 dB), nominal
10 Hz to 150 kHz
fixed lowpass filter
150 kHz to 30 MHz
35 MHz, fixed bandpass filter
30 MHz to 80 MHz
94 MHz, fixed bandpass filter
80 MHz to 130 MHz
94 MHz, fixed bandpass filter
130 MHz to 180 MHz
91 MHz, fixed bandpass filter
180 MHz to 230 MHz
105 MHz, fixed bandpass filter
230 MHz to 300 MHz
110 MHz, fixed bandpass filter
300 MHz to 425 MHz
195 MHz, fixed bandpass filter
425 MHz to 570 MHz
200 MHz, fixed bandpass filter
570 MHz to 715 MHz
210 MHz, fixed bandpass filter
715 MHz to 860 MHz
200 MHz, fixed bandpass filter
860 MHz to 1005 MHz
200 MHz, fixed bandpass filter
1005 MHz to 1750 MHz
fixed highpass filter
1750 MHz to 2850 MHz
fixed highpass filter
2850 MHz to 4850 MHz
fixed highpass filter
4850 MHz to 7000 MHz
fixed highpass filter
Preamplifier
switchable
Location
in the signal path between preselection
and 1st mixer
Frequency range
1 kHz to 7 GHz
Gain
20 dB (nom.)
RF preamplifier (R&S
®
FSV-B22 option)
Preamplifier
Availability
instruments without R&S
®
ESRP-B2 option receiver, analyzer
instruments with R&S
®
ESRP-B2 option
analyzer with preselection = off only
Frequency range
100 kHz to 7 GHz
Gain
20 dB (nom.)
IF and resolution bandwidths
IF and sweep filters
Resolution bandwidths (–3 dB)
receiver mode or analyzer mode,
span ≥ 10 Hz
10 Hz to 10 MHz in 1/2/3/5 sequence
analyzer mode, span = 0 Hz
20 MHz, 28 MHz, 40 MHz additionally
Bandwidth uncertainty
< 3 %
Shape factor 60 dB:3 dB
< 5
EMI bandwidths (–6 dB)
standard
200 Hz, 9 kHz, 120 kHz, 1 MHz
with R&S
®
ESRP-B29 option
10 Hz, 100 Hz, 1 kHz, 10 kHz, 100 kHz
additionally
Bandwidth uncertainty
< 3 %
Shape factor 60 dB:6 dB
< 4
FFT filters (analyzer mode)
Resolution bandwidths (–3 dB)
span ≥ 10 Hz
10 Hz to 300 kHz in 1/2/3/5 sequence
Bandwidth uncertainty
< 3 % (nom.)
Shape factor 60 dB:3 dB
< 5 (nom.)