Atec Rohde-Schwarz-ESRP Series User Manual
Page 10
Version 01.01, January 2013
10
Rohde & Schwarz R&S
®
ESRP EMI Test Receiver
Level display (analyzer mode)
Logarithmic level axis
1 dB to 200 dB, in steps of 1/2/5
Linear level axis
10 % of reference level per level division,
10 divisions or logarithmic scaling
Number of traces
6
Trace detector
max. peak, min. peak, auto peak (normal),
sample, RMS, average, quasi-peak,
CISPR-average, RMS-average
Trace functions
clear/write, max. hold, min. hold, average,
view
Setting range of reference level
–130 dBm to (–10 dBm + RF attenuation
– RF preamplifier gain), in steps of
0.01 dB
Units of level axis
logarithmic level display
dBm, dBµV, dBmV, dBµA, dBpW
linear level display
µV, mV, µA, mA, pW, nW
Level display (receiver mode)
Level display
analog
bargraph display, separately for each
detector
digital
numeric; 0.01 dB resolution
Detectors
max. 4 selectable
max. peak, min. peak, RMS, average,
quasi-peak, CISPR-average,
RMS-average
Units of level axis
dBm, dBµV, dBmV, dBµA, dBpW, dBpT
RF spectrum
Logarithmic level axis
10 dB to 200 dB, in steps of 10
Frequency axis
linear or logarithmic
Number of traces
6
Detectors
normal scan
max. peak, min. peak, RMS, average,
quasi-peak, CISPR-average,
RMS-average
time domain scan
8
max. peak, min. peak, average,
quasi-peak, CISPR-average,
RMS-average
Spectrogram display (analyzer mode)
Result display
color-graded bitmap
Spectrogram bitmap color depth
240 colors
Dynamic range covered by bitmap colors
selectable, up to 200 dB (nom.)
History depth
max. 100 000 frames
Recording mode
single trace, continuous, frame count
Trace detector
max. peak, min. peak, sample, RMS,
average
Number of markers
16
Marker readout
frequency, time/frame number, level
8
Requires R&S
®
ESRP-K53 option.