Atec Rohde-Schwarz-ESRP Series User Manual
Page 8
Version 01.01, January 2013
8
Rohde & Schwarz R&S
®
ESRP EMI Test Receiver
Second harmonic intercept (SHI)
RF attenuation = 0 dB, level = –10 dBm, preselection = off
5
,
with R&S
®
FSV-B22 option: RF preamplifier = off
100 MHz < f
in
≤ 3.5 GHz
45 dBm (nom.)
with R&S
®
ESRP-B2 option,
RF attenuation = 0 dB, level = –15 dBm, preselection = on
6
, preamplifier = off
100 MHz < f
in
≤ 3.5 GHz
50 dBm (nom.)
with R&S
®
ESRP-B2 option,
RF attenuation = 0 dB, level = –10 dBm, preselection = on
6
, preamplifier = on
100 MHz < f
in
≤ 3.5 GHz
35 dBm (nom.)
with R&S
®
FSV-B22 option,
preselection = off
5
, RF preamplifier = on, RF attenuation = 0 dB, level = –40 dBm
100 MHz < f
in
≤ 3.5 GHz
25 dBm (nom.)
Displayed average noise level
(analyzer mode)
RF attenuation = 0 dB, preselection = off/on
7
, preamplifier = off, termination = 50 Ω,
log. scaling, normalized to 1 Hz RBW, RBW = 1 kHz, VBW = 3 kHz, zero span, sweep
time = 50 ms, sample detector, trace average, sweep count = 20, mean marker
R&S
®
ESRP3, R&S
®
ESRP7
9 kHz ≤ f < 100 kHz
< –130 dBm,–140 dBm typ.
100 kHz ≤ f < 1 MHz
< –145 dBm,–150 dBm typ.
1 MHz ≤ f < 1 GHz
< –152 dBm,–155 dBm typ.
1 GHz ≤ f < 3.6 GHz
< –150 dBm,–151 dBm typ.
R&S
®
ESRP7
3.6 GHz ≤ f < 6 GHz
< –148 dBm,–151 dBm typ.
6 GHz ≤ f ≤ 7 GHz
< –146 dBm,–149 dBm typ.
with R&S
®
ESRP-B29 option, RF attenuation = 0 dB, preselection = off/on
7
,
preamplifier = off, termination = 50 Ω, log. scaling, normalized to 1 Hz RBW,
RBW = 10 Hz, VBW = 10 Hz, zero span, sweep time = 500 ms, sample detector,
trace average, sweep count = 20, mean marker
R&S
®
ESRP3, R&S
®
ESRP7
10 Hz
< –90 dBm,–100 dBm typ.
20 Hz
< –100 dBm,–110 dBm typ.
100 Hz
< –110 dBm,–120 dBm typ.
1 kHz
< –120 dBm,–130 dBm typ.
with R&S
®
ESRP-B2 option,
RF attenuation = 0 dB, preselection = on, preamplifier = on, termination = 50 Ω,
log. scaling, normalized to 1 Hz RBW, RBW = 1 kHz, VBW = 3 kHz, zero span,
sweep time = 50 ms, sample detector, trace average, sweep count = 20, mean marker
R&S
®
ESRP3, R&S
®
ESRP7
9 kHz ≤ f < 100 kHz
< –150 dBm,–155 dBm typ.
100 kHz ≤ f < 1 MHz
< –155 dBm,–160 dBm typ.
1 MHz ≤ f < 1 GHz
< –165 dBm,–168 dBm typ.
1 GHz ≤ f < 3.6 GHz
< –162 dBm,–165 dBm typ.
R&S
®
ESRP7
3.6 GHz ≤ f < 6 GHz
< –160 dBm,–163 dBm typ.
6 GHz ≤ f ≤ 7 GHz
< –158 dBm, typ. –161 dBm
with R&S
®
ESRP-B2 and R&S
®
ESRP-B29 option,
RF attenuation = 0 dB, preselection = on, preamplifier = on, termination = 50 Ω,
log. scaling, normalized to 10 Hz RBW, RBW = 10 Hz, VBW = 5 Hz, zero span,
sweep time = 500 ms, sample detector, trace average, sweep count = 20, mean marker
R&S
®
ESRP3, R&S
®
ESRP7
1 kHz
< –140 dBm,–150 dBm typ.
with R&S
®
FSV-B22 option,
RF attenuation = 0 dB, preselection = off
7
, RF preamplifier = on, termination = 50 Ω,
log. scaling, normalized to 1 Hz RBW, RBW = 1 kHz, VBW = 3 kHz, zero span,
sweep time = 50 ms, sample detector, trace average, sweep count = 20, mean marker
R&S
®
ESRP3, R&S
®
ESRP7
100 kHz ≤ f < 1 MHz
< –150 dBm,–155 dBm typ.
1 MHz ≤ f < 1 GHz
< –162 dBm,–165 dBm typ.
1 GHz ≤ f < 3.6 GHz
< –160 dBm,–163 dBm typ.
R&S
®
ESRP7
3.6 GHz ≤ f < 6 GHz
< –158 dBm,–161 dBm typ.
6 GHz ≤ f ≤ 7 GHz
< –156 dBm, typ. –159 dBm
7
With R&S
®
ESRP-B2 option. Preselection = off is only available in analyzer mode. In receiver mode preselection is permanently on.