Level – Atec Rohde-Schwarz-ESRP Series User Manual
Page 7
Version 01.01, January 2013
Rohde & Schwarz R&S
®
ESRP EMI Test Receiver
7
Channel filters (analyzer mode)
Bandwidths (–3 dB)
standard
(RRC = root raised cosine)
100/200/300/500 Hz
1/1.5/2/2.4/2.7/3/3.4/4/4.5/5/6/8.5/9/10/
12.5/14/15/16/18 (RRC)/20/21/24.3 (RRC)
/25/30/50/100/150/192/200/300/500 kHz
1/1.228/1.28 (RRC)/1.5/2/3/3.84 (RRC)/
4.096 (RRC)/5/10/20/28/40 MHz
Bandwidth accuracy
< 2 % (nom.)
Shape factor 60 dB:3 dB
< 2 (nom.)
Video bandwidths (analyzer mode)
1 Hz to 10 MHz in 1/2/3/5 sequence,
20 MHz, 28 MHz, 40 MHz
Level
Display range
displayed noise floor up to +130 dBm
Max. input level
AC coupled
50 V
DC coupled
0 V
CW RF power
RF attenuation = 0 dB
RF preamplifier = off
20 dBm (= 0.1 W)
RF preamplifier = on
13 dBm (= 0.02 W)
RF attenuation ≥ 10 dB
RF preamplifier = off
30 dBm (= 1 W)
RF preamplifier = on
23 dBm (= 0.2 W)
Pulse spectral density
RF attenuation ≥ 0 dB, preselection = on
4
,
RF preamplifier = off
97 dB µV/MHz
Max. pulse voltage
RF attenuation ≥ 10 dB
150 V
Max. pulse energy
RF attenuation ≥ 10 dB, 10 µs
1 mWs
Intermodulation
1 dB compression of input mixer
RF attenuation = 0 dB, preselection and
preamplifier = off
5
+3 dBm (nom.)
Third-order intercept point (TOI)
RF attenuation = 0 dB, level = 2 × –15 dBm, ∆f > 5 × RBW or 10 kHz, whichever is
larger, preselection = off
5
, with R&S
®
FSV-B22 option: RF preamplifier = off
10 MHz ≤ f
in
< 100 MHz
> 12 dBm, 15 dBm typ.
100 MHz ≤ f
in
< 3.6 GHz
> 13 dBm, 16 dBm typ.
3.6 GHz ≤ f
in
≤ 7 GHz
> 15 dBm, 18 dBm typ.
with R&S
®
ESRP-B2 option,
preselection = on
6
, preamplifier = off, RF attenuation = 0 dB, level = 2 × –20 dBm,
∆f > 5 × RBW or 10 kHz, whichever is larger
10 MHz ≤ f
in
< 100 MHz
> 5 dBm, 8 dBm typ.
100 MHz ≤ f
in
< 4.5 GHz
> 8 dBm, 11 dBm typ.
4.5 GHz ≤ f
in
≤ 7 GHz
> 5 dBm, 8 dBm typ.
with R&S
®
ESRP-B2 option,
preselection = on
6
, preamplifier = on, RF attenuation = 0 dB, level = 2 × –45 dBm,
∆f > 5 × RBW or 10 kHz, whichever is larger
10 MHz ≤ f
in
< 100 MHz
> –16 dBm,–13 dBm typ.
100 MHz ≤ f
in
< 3.6 GHz
> –14 dBm,–11 dBm typ.
3.6 GHz ≤ f
in
≤ 7 GHz
> –10 dBm,–7 dBm typ.
with R&S
®
FSV-B22 option,
preselection = off
5
, RF preamplifier = on, RF attenuation = 0 dB, level = 2 × –45 dBm,
∆f > 5 × RBW or 10 kHz, whichever is larger
10 MHz ≤ f
in
< 100 MHz
–3 dBm (nom.)
100 MHz ≤ f
in
< 3.6 GHz
–2 dBm (nom.)
3.6 GHz ≤ f
in
≤ 7 GHz
0 dBm (nom.)
4
With R&S
®
ESRP-B2 option.
5
With R&S
®
ESRP-B2 option. Preselection = off is only available in analyzer mode. In receiver mode the preselection is permanently on.
6
With R&S
®
ESRP-B2 option. Default setting in receiver mode.