beautypg.com

General data – Atec Rohde-Schwarz-ESRP Series User Manual

Page 15

background image

Version 01.01, January 2013

Rohde & Schwarz R&S

®

ESRP EMI Test Receiver

15

General data

Display

21 cm LC TFT color display (8.4")

Resolution

800 × 600 pixel (SVGA resolution)

Pixel failure rate

< 1 × 10

–5

Data storage
Internal standard hard

disk

≥ 40 Gbyte

with R&S

®

ESRP-B18 option

solid state disk ≥ 8 Gbyte

External

supports USB 2.0 compatible memory
devices

Temperature
Temperature

operating temperature range

+5 °C to +40 °C

permissible temperature range

0 °C to +50 °C

storage temperature range

–40 °C to +70 °C

Climatic loading

+40 °C at 90 % rel. humidity,
in line with EN 60068-2-30

Mechanical resistance
Vibration

sinusoidal

5 Hz to 150 Hz, max. 2 g at 55 Hz;
0.5 g from 55 Hz to 150 Hz;
in line with EN 60068-2-6

random

10 Hz to 130 Hz, acceleration
1.2 g (RMS), in line with EN 60068-2-64

Shock

40 g shock spectrum,
in line with MIL-T-28800F, class 3,
MIL-STD-810E, method 516.4, procedure I

EMC

in line with EMC Directive 2004/108/EC
including:
IEC/EN 61326-1

12,

13

IEC/EN 61326-2-1
CISPR 11/EN 55011

12

IEC/EN 61000-3-2
IEC/EN 61000-3-3

Recommended calibration interval

1

year

Power supply
AC input voltage range

100 V to 240 V, ±10 % (nom.)

AC supply frequency

50 Hz to 400 Hz, +10 %/–6 % (nom.)

Max. input current

3 A (100 V) to 1.25 A (240 V) (nom.)

Power consumption

100 W, max. 180 W with all options
(meas.)

Safety

in line with IEC 61010-1, EN 61010-1,
CAN/CSA-C22.2 No. 61010-1-04,
UL 61010-1

Test mark

VDE-GS,

C

CSA

US

Dimensions and weight
Dimensions (nom.)

W × H × D

412 mm × 197 mm × 417 mm
(16.22 in × 7.76 in × 16.42 in)

Net weight without options (nom.)

9.5 kg (20.94 lb)

12

Emission limits for class B equipment.

13