beautypg.com

Specifications, Frequency – Atec Rohde-Schwarz-ESRP Series User Manual

Page 4

background image

Version 01.01, January 2013

4

Rohde & Schwarz R&S

®

ESRP EMI Test Receiver

Specifications

Operating modes

EMI test receiver
spectrum analyzer

Frequency

Frequency range

R&S

®

ESRP3

AC coupled

10 MHz to 3.6 GHz

DC coupled

9 kHz to 3.6 GHz

R&S

®

ESRP7

AC coupled

10 MHz to 7 GHz

DC coupled

9 kHz to 7 GHz

with R&S

®

ESRP-B29 option, DC coupled

10 Hz to max. frequency

Frequency resolution

receiver mode

0.1 Hz

analyzer mode

0.01 Hz

Reference frequency, internal
Accuracy

±((time since last adjustment × aging rate)
+ temperature drift + calibration accuracy)

Aging per year

standard

±1 × 10

–6

with R&S

®

FSV-B4 option

±1 × 10

–7

Temperature drift (+5 °C to +45 °C)

standard

±1 × 10

–6

with R&S

®

FSV-B4 option, model .02

±1 × 10

–7

with R&S

®

FSV-B4 option, model .03

±1 × 10

–8

Max. initial calibration accuracy

standard

±5 × 10

–7

with R&S

®

FSV-B4 option

±5 × 10

–8

Frequency readout (analyzer mode)
Marker resolution

1 Hz

Uncertainty

±(marker frequency × reference accuracy
+ 10 % × resolution bandwidth +
½ (span/(sweep points – 1)) + 1 Hz)

Number of sweep (trace) points

default value

691

range

spectrum analyzer

101 to 32 001

EMI measurement

101 to 200 001

Marker tuning frequency step size

marker step size = sweep points

span/(sweep points – 1)

marker step size = standard

span/(default sweep points – 1)

Frequency counter resolution

0.001 Hz

Count accuracy

±(frequency × reference accuracy +
½ (last digit))

Display range for frequency axis

0 Hz, 10 Hz to max. frequency

Resolution

0.1

Hz

Max. span deviation

±0.1 %

Receiver scan
Scan

scan with max. 10 subranges with different
settings

Scan modes

normal scan, time domain scan

1

Measurement time

normal scan, per frequency

50 µs to 100 s

time domain scan, per subrange

1

50 µs to 100 s

Number of trace points

up to 4 000 000

Frequency step size

normal scan

min. 1 Hz

time domain scan

1

0.25 × resolution bandwidth

Time domain scan

1

Frequency segment processed in parallel

RBW = 200 Hz

0.66 MHz

RBW = 9 kHz

30 MHz

RBW = 120 kHz

24.6 MHz

RBW = 1 MHz

25.6 MHz

FFT overlap factor

≥ 93 %

1

Requires R&S

®

ESRP-K53 option.