MPRT

Megger Protective Relay Test System

7

With three currents in parallel:
Output Current

Power

Max V

12 Amperes

600 VA

50 Vrms

22.5Amperes

600 VA

26.7 Vrms

45 Amperes

600 VA

13.4 Vrms

With four currents in parallel:
Output Current

Power

Max V

16 Amperes

800 VA

50 Vrms

30 Amperes

800 VA

26.7 Vrms

60 Amperes

800 VA

13.4 Vrms

Battery Simulator

The battery simulator provides the following DC output:
24, 48, 125 or 250 Volts at 60 Watts. Voltage output is controlled
via the Touch-View Interface, or through AVTS software.

Metering

Measured output quantities such as AC Amperes, AC Volts, DC
Volts or DC Amperes, and Time may be simultaneously displayed
on the large, variable contrast, color LCD touch screen. The
memory feature of the metering provides fast and accurate preset
of test values. The AC and DC outputs display the approximate
voltage/current output prior to initiation of the outputs. This
provides a fast, easy method for preset of outputs. Other values
that may be displayed, depending on which test screen is in view,
are phase angle, frequency, Ohms, Watts, VARS and Power Factor.
All Accuracies stated are from 10 to 100% of the range at 50/60 Hz.

AC Voltage Amplitude

Accuracy:

±0.05% typical, 0.1% guaranteed.

Resolution: .0001/.001/.01
Measurements: True RMS

AC Current Amplitude

Accuracy: ±0.05% typical, 0.1% or ±20 mA, which ever is greater,
guaranteed.
Resolution: .0001/.001/.01
Measurements: True RMS

DC Voltage Amplitude

Accuracy: ±0.1% typical, 0.25 % guaranteed.
Resolution: .0001/.001/.01
Measurements: True RMS

DC Current Amplitude

Accuracy: ±0.1% typical, 0.25 % or ±20 mA, which ever is greater,
guaranteed.
Resolution: .0001/.001/.01
Measurements: True RMS

Convertible Source in AC Current Mode

Accuracy: ±0.05% typical, 0.1 % or ±12.5 mA,
which ever is greater, guaranteed.
Resolution: .0001
Measurements: True RMS

Phase Angle

Ranges 0.01 to 359.99 degrees, Counter Clock Wise,
or Clock Wise rotation, or ±0.01 to ±180.00 degrees
Accuracy: ± 0.02° typical
± 0.25° at 50/60 Hz max

Frequency:
The output modules provide a variable frequency output with the
following ranges and accuracy.

Ranges
DC
0.001 to 1000.000 Hz

Output amplifiers can provide transient signals with a range of
DC to 10 kHz for transient playback.

Resolution: .001 Hz
Frequency Accuracy:
2.5 ppm typical
25 ppm

0° to 50° C, at 50/60 Hz

Total Harmonic Distortion:
Less than 0.1% typical,
2% maximum at 50/60 Hz.

Power Curve for Model 8415

With two currents in series, the compliance voltage doubles to
provide 4.0 Amperes at 100 Volts.

AC Voltage Output
The following specifications cover both the Model 8430 and 8415
modules. Outputs are rated with the following Ranges:

Model 8430

Per phase:
Output Volts

Power

Max I

30 Volts

150 VA

5 Amps

150 Volts

150 VA

1 Amp

300 Volts

150 VA

0.5 A

DC

150 Watts

With two voltages in series:
Output Volts

Power

Max I

60 Volts

300 VA

5 Amps

300 Volts

300 VA

1 A

600 Volts

300 VA

0.5 A

With three converted sources in parallel:
Output Current

Power

Max V

15Amperes

450 VA

30 Vrms

Model 8415

Per phase:
Output Volts

Power

Max I

30 Volts

150 VA

5 Amps

150 Volts

150 VA

1 Amp

DC

150 Watts

With two voltages in series:
Output Volts

Power

Max I

60 Volts

300 VA

5 Amps

300 Volts

300 VA

1 A

With three converted sources in parallel:
Output Current

Power

Max V

15Amperes

450 VA

30 Vrms