Audio Analyzer UPL

17

Modulation factor (MOD DIST)
Measurement method

selective to DIN IEC 268-3

Frequency range

lower frequency 30 Hz to 2700 Hz
upper frequency 8 x LF to 100 kHz

1)

Accuracy

±

0.50 dB

Inherent distortion

2)

Upper frequency 4 kHz to 15 kHz

<

–96 dB (–90 dB), typ. –103 dB

15 kHz to 20 kHz

<

–96 dB (–85 dB)

Spectrum

bar chart showing signal and distortion

Difference frequency distortion (DFD)
Measurement method

selective to DIN IEC 268-3 or 118

Frequeny range

difference frequency 80 Hz to 2 kHz
center frequency 200 Hz to 100 kHz

3)

Accuracy

±

0.50 dB, center frequency

<

20 kHz

Inherent distortion

4)

DFD d

2

<

–112 dB, typ. –125 dB

DFD d

3

<

–96 dB, typ. –105 dB

Spectrum

bar chart showing signal and distortion

Wow and flutter

with analyzer ANLG 22 kHz only

Measurement method

DIN/IEC, NAB, JIS,
2-sigma to IEC-386

Weighting filter

OFF

highpass 0.5 Hz, bandwidth 200 Hz

ON

bandpass 4 Hz to IEC-386

Accuracy

±

3%

Inherent noise

<

0.0005% weighted

<

0.001% unweighted

Spectrum

post-FFT of demodulated signal

Time domain display (WAVEFORM)
Trigger

rising/falling edge

Trigger level

–200 V to +200 V, interpolated between
samples

Trace length

max. 7424 points

Standard mode

1- to 32-fold interpolation

Compressed mode

2- to 1024-fold compression
(envelope for AGC measurement), with
analyzer ANLG 22 kHz only

Frequency

5)

Frequency range

20 Hz to 110 kHz

Accuracy

±50 ppm

Phase

5)

with analyzer 22 kHz only

Frequency range

20 Hz to 20 kHz

Accuracy

±0.5°

Group delay

5)

with analyzer 22 kHz only

Frequency range

20 Hz to 20 kHz

Accuracy in seconds

∆ϕ

/(

f x360), where

∆ϕ

= phase accu-

racy in °,

f = frequency step

Polarity test
Measurement

polarity of unsymmetrical input signal

Display

+POL, –POL

1)

For upper frequency

>

20 kHz, the bottom limit of lower frequency is reduced.

2)

Input voltage

>

200 mV, typical values apply between 0.5 V and 3.5 V.

Lower frequency

>

200 Hz, values in ( ) for lower frequency

<

200 Hz.

Dynamic mode precision; level ratio LF:UF = 4:1.

3)

For center frequencies

>

20 kHz the bottom limit of the difference frequency is

reduced.

4)

Input voltage >200 mV, typical values apply between 0.5 V and 3.5 V,
dynamic mode precision (at DFD d2), center frequency 7 kHz to 20 kHz.

5)

With measurement functions RMS, FFT and THD+N only, accuracy applies to
8k FFT with zoom factor 2, Rife-Vincent-2 window; S/N ratio >70 dB.

Analog generators

An 18-bit

∆Σ

D/A converter is used for analog signal generation. The charac-

teristics of the basic generator can be improved and extended with a low-
distortion RC oscillator (Low Distortion Generator UPL-B1):
– sine with reduced distortion
– frequency range up to 110 kHz

Outputs

XLR connectors, 2 channels, floating, balanced/unbalanced switchable, short-
circuit-proof; max. current

<

120 mA with external feed

Balanced
Voltage

0.1 mV to 20 V (rms, sine, open-circuit)

Crosstalk attenuation

>

115 dB, frequency

<

20 kHz

Source impedance

typ. 10

,

200

(150

with UPL-U3)

±

0.5%, 600

Ω ±

0.5%

Load impedance

>

400

(incl. source impedance)

Output balance

>

75 dB at 1 kHz,

>

60 dB at 20 kHz

Unbalanced
Voltage

0.1 mV to 10 V (rms, sine, open-circuit)

Crosstalk attenuation

>

115 dB, frequency

<

20 kHz

Source impedance

5

Load impedance

>200

Signals

Sine
Frequency range

2 Hz to 21.75 kHz

Frequency accuracy

±

50 ppm

Level accuracy

±

0.1 dB at 1 kHz

Frequency response (ref. to 1 kHz)

20 Hz to 20 kHz

±

0.05 dB

Inherent distortion THD+N

Measurement bandwidth
20 Hz to 22 kHz

<

–94 dB, typ. –98 dB

20 Hz to 100 kHz

<

–86 dB

Sweep parameters

frequency, level

Sine (with low distortion generator option)
Frequency range

10 Hz to 110 kHz

Frequency accuracy

±

0.5% at 15°C to 30

°

C

±

0.75% at 5°C to 45

°

C

Level accuracy

±

0.1 dB at 1 kHz

Frequency response (ref. to1 kHz)

20 Hz to 20 kHz

±

0.05 dB

10 Hz to 110 kHz

±

0.1 dB

Harmonics typ.

<

–115 dB (

<

–120 dB at 1 kHz),

measurement bandwidth 20 Hz to
20 kHz, voltage 1V to 5 V

Inherent distortion (THD)

Fundamental 1 kHz, 1 V to 10 V

<

–120 dB typ.

20 Hz to 7 kHz

<

–105 dB

7 kHz to 20 kHz

<

–100 dB

Inherent distortion (THD+N)

6)

Meas. bandw.

Fundamental 1 kHz, 2.5 V

–110 dB typ.

22 kHz

20 Hz to 20 kHz

<

–100 dB +2

µ

V

22 kHz

20 Hz to 20 kHz

<

–88 dB +5

µ

V

100 kHz

Sweep parameters

frequency, level

6)

Total inherent distortion of analyzer and generator, analyzer with dynamic
mode precision.