Audio Analyzer UPL

Specifications

Data without tolerances are typical values.

Analog analyzers

For analog measurements two analyzers with different bandwidths, specifica-
tions and measurement functions are available:
Analyzer

Frequency range

ANLG 22 kHz

DC/10 Hz to 21.90 kHz

1)

ANLG 110 kHz

DC/20 Hz to110 kHz

1)

Level measurements (rms)

Accuracy at 1 kHz

±

0.05 dB

Frequency response ref. to 1 kHz)
20 Hz to 22 kHz

±

0.03 dB, typ. 0.003 dB (V

in

<3 V)

10 Hz to 20 Hz

±

0.1 dB

22 kHz to 50 kHz

±

0.1 dB

50 kHz to 110 kHz

±

0.2 dB

Inputs

XLR connectors

2 channels, balanced (unbalanced
measurements possible with XLR/BNC
Adapter UPL-Z1), floating/grounded
and AC/DC coupling switchable

Voltage range

0.1

µ

V to 110 V (rms, sine)

Measurement ranges

18 mV to 100 V, in steps of 5 dB

Input impedance

100 k

Ω ±

1% shunted by 120 pF, each

pin against ground
300

, 600

,

±

0.5% each, P

max

1 W

Crosstalk attenuation

>

120 dB, frequency

<

22 kHz, 600

Common-mode rejection (V

in

<

3 V)

>

100 dB at 50 Hz,

>

86 dB at 1 kHz,

>

80 dB at 16 kHz

Generator output

each input channel switchable to the
other output channel, input impedance:
balanced 200 k

, unbalanced

100 k

Typical frequency response, measured with internal generator/analyzer at
analog interfaces

Measurement functions

RMS value, wideband
Accuracy

Measurement speed

AUTO

±

0.05 dB at 1 kHz, sine

AUTO FAST

±

0.1 dB additional error

Integration time

AUTO FAST/AUTO

4.2 ms/42 ms, at least 1 cycle

VALUE

1 ms to 10 s

GEN TRACK

2.1 ms, at least 1 cycle

Noise (600

)

with A filter

1

µ

V

with CCIR unweighting filter

<

2

µ

V, 1.6

µ

V typ. (ANLG 22 kHz)

Filter

weighting filters and user-definable fil-
ters, up to 3 filters can be combined,
analog notch filter in addition (expan-
sion of dynamic range by up to 30 dB)

Spectrum

post-FFT of filtered signal

1)

DC/AC coupling.

RMS value, selective
Bandwidth (–0.1 dB)

1%, 3%, 1/12 octave, 1/3 octave and
user-selectable fixed bandwidth,
minimum bandwidth 20 Hz

Selectivity

100 dB (80 dB) with analyzer ANLG
22 kHz (110 kHz) bandpass or band-
stop filter, 8th order elliptical filter, ana-
log notch filter in addition

Frequency setting

– automatic to input signal
– coupled to generator
– fixed through entered value
– sweep in selectable range

Accuracy

±

0.2 dB + ripple of filters

Peak value

with analyzer ANLG 22 kHz only

Measurement

peak max, peak min, peak-to-peak,
peak absolute

Accuracy

±

0.2 dB at 1 kHz

Interval

20 ms to 10 s

Filter

2)

weighting filters and user-definable
filters, up to 3 filters can be combined

Quasi-peak

with analyzer ANLG 22 kHz only

Measurement, accuracy

to CCIR 468-4

Noise (600

)

<

8

µ

V with CCIR weighting filter

Filter

2)

weighting filters and user-definable
filters, up to 3 filters can be combined,
analog notch filter in addition

DC voltage
Voltage range

0 V to

±

110 V

Accuracy

±

(1% of measured value + 0.1% of

measurement range)

Measurement ranges

100 mV to 100 V, in steps of 10 dB

S/N measurement routine

available for measurement functions
– rms, wideband
– peak
– quasi-peak
indication of S/N ratio in dB,
no post-FFT

FFT analysis

see FFT analyzer section

Total harmonic distortion (THD)
Fundamental

10 Hz to 22 kHz

Frequency tuning

automatic to input or generator signal
or fixed through entered value

Weighted harmonics

any combination of d

2

to d

9

,

up to 110 kHz

Accuracy

Harmonics

<

50 kHz

±

0.5 dB

<

110 kHz

±

0.7 dB

Inherent distortion

3)4)

Analyzer ANLG 22 kHz
Fundamental

20 Hz to 10.95 kHz

<

–110 dB, typ. –115 dB

10 Hz to 20 Hz

<

–100 dB

Analyzer ANLG 110 kHz
Fundamental

50 Hz to 20 kHz

<

–100 dB, typ. –105 dB

Spectrum

bar chart showing signal and distortion

THD+N and SINAD
Fundamental

10 Hz to 22 kHz

Frequency tuning

automatic to input or generator signal
or fixed through entered value

Input voltage

typ.

>

100

µ

V with automatic tuning

Bandwidth

upper and lower frequency limit se-
lectable, one weighting filter in addi-
tion

Accuracy
Bandwidth

<

50 kHz

±

0.5 dB

<

100 kHz

±

0.7 dB

Inherent distortion

3)

Analyzer ANLG 22 kHz

Bandwidth 20 Hz to 21.90 kHz typ. –110 dB at 1 kHz, 2.5 V

<

–105 dB +2

µ

V

5)

typ. –108 dB +1.5

µ

V

Analyzer ANLG 110 kHz

Bandwidth 20 Hz to 22 kHz

<

–95 dB + 2.5

µ

V, typ. –100 dB +1.75

µ

V

20 Hz to 110 kHz

<

–88 dB + 5

µ

V, typ. –95 dB + 3.5

µ

V

Spectrum

post-FFT of filtered signal

2)

With UPL-B29 only in base rate mode.

3)

Total inherent distortion of analyzer and generator (with option
UPL-B1), analyzer with dynamic mode precision.

4)

>3.5 V: typ. 3 dB less; <0.5 V: sensitivity reduced by inherent noise
(typ. 0.25/1.25

µ

V with analyzers 22/110 kHz).

5)

At full-scale level of measurement range (<–100 dB + 2

µ

V with auto range),

<–100 dB for input voltage >3.5 V.