beautypg.com

Analog generators, Outputs, Signals – Atec Rohde-Schwarz-UPV User Manual

Page 8

background image

Version 03.00, August 2011

8

Rohde & Schwarz R&S

®

UPV Audio Analyzer

Analog generators

Outputs

XLR connectors (pin 1 not connected), 2 channels, floating/grounded selectable, balanced/unbalanced selectable, short-circuit-proof;
max. current

< 120 mA with external feed.

Balanced
Voltage

RMS, sine, open circuit

0.1 mV to 20 V

Crosstalk attenuation

frequency < 20 kHz

> 115 dB

typ.

10

200

Ω (150 Ω with R&S

®

UPV-U1) ±

0.5 %

Source impedance

600

± 0.5 %

Load impedance

incl. source impedance

> 400 Ω

at 1 kHz

> 75 dB

Output balance

at 20 kHz

> 60 dB

Unbalanced
Voltage

RMS, sine, open circuit

0.1 mV to 10 V

Crosstalk attenuation

frequency < 20 kHz

> 115 dB

Source impedance

typ. 5

Load impedance

> 200 Ω

Signals

Sine
Frequency range

0.1 Hz to 80 kHz

Frequency error

±10 ppm

Level error

at 1 kHz

±0.05 dB

Frequency response (referenced to 1 kHz) 20 Hz to 20/70/80 kHz

±0.01 dB/±0.05dB/±0.1 dB

level < 3 V
fundamental meas.

bandwidth

20 Hz to 20 kHz

22 kHz


< –103 dB + 2.5 µ

V, typ. –107 dB

Inherent distortion (THD+N)

20 Hz to 20 kHz

80 kHz

< –90 dB + 5 µ

V

Sweep parameters

frequency, level

Sine (with R&S

®

UPV-B1 low distortion generator option)

Frequency range

10 Hz to 185 kHz

10 Hz to 100 kHz

±0.5 %

Frequency error

100 kHz to 185 kHz

±0.75 %

Level error

at 1 kHz

±0.05 dB

20 Hz to 20 kHz

±0.01 dB

10 Hz to 100 kHz

±0.05 dB

100 kHz to 150 kHz

±0.15 dB

Frequency response (referenced to 1 kHz)

150 kHz to 185 kHz

±0.25 dB

Harmonics

measurement bandwidth 20 Hz to 20 kHz,
voltage 1 V to 5 V

typ. < –115 dB (< –120 dB at 1 kHz)

fundamental
1 kHz, 1 V to 10 V

typ. < –120 dB

20 Hz to 7 kHz

< –110 dB, typ. –115 dB

7 kHz to 20 kHz

< –105 dB, typ. –110 dB

20 kHz to 50 kHz

< –88 dB

Inherent distortion (THD)

50 kHz to 100 kHz

< –80 dB

fundamental meas.

bandwidth

1 kHz, 2.5 V

22 kHz


typ. –110 dB

20 Hz to 20 kHz

22 kHz

< –100 dB + 2 µV

Inherent distortion (THD+N)

8

20 Hz to 20 kHz

100 kHz

< –88 dB + 5 µV

Sweep parameters

frequency, level


8

Total inherent distortion of analyzer and generator, analyzer with dynamic mode precision.