beautypg.com

Remote control (r&s, Upv-k4 option), Extended analysis functions (r&s – Atec Rohde-Schwarz-UPV User Manual

Page 24: Upv-k6 option), Pesq, Measurement (r&s, Upv-k61 option), Peaq, Upv-k62 option), Polqa

background image

Version 03.00, August 2011

24

Rohde & Schwarz R&S

®

UPV Audio Analyzer

Remote control (R&S

®

UPV-K4 option)

Enables remote control via IEC 625-2 (IEEE 488), LAN and USB.

Commands largely compliant with SCPI.

Extended analysis functions (R&S

®

UPV-K6 option)

Rub & buzz measurement

simultaneous measurement of frequency
response, rub & buzz and polarity

Frequency range

20 Hz to 80 kHz

Tracking highpass filter

2 to 20 times fundamental frequency

Lower/upper frequency limit

selectable

1/n octave analysis
Frequency range

20 Hz to 20 kHz

at center frequency

±0.2 dB

Level error

20 Hz to 20 kHz

±1.0 dB (EN 61260, class 0)

Undersample FFT

FFT resolution is improved while reducing
the measurement bandwidth

Undersampling factor

up to 1024

Highest resolution

bandwidth 0 Hz to 23 Hz

0.18 mHz

PESQ

®

measurement (R&S

®

UPV-K61 option)

16

Perceptual evaluation of speech quality

in line with ITU-T recommendation P.862,
862.1 and 862.2

Numeric results

PESQ score,
MOS-LQO narrowband and wideband,
average delay

Graphic displays (versus time)

PESQ score, MOS-LQO, delay, dropouts,
reference signal and degraded signal

PEAQ

®

measurement (R&S

®

UPV-K62 option)

16

Perceptual evaluation of audio quality

in line with ITU-R recommendation
BS.1387

Numeric results

ODG (objective difference grade),
DI (distortion index),
average delay

R&S

®

UPV-B3 option required for dual channel analog signals.

POLQA

®

measurement (R&S

®

UPV-K63 option)

16

Perceptual objective listening quality
analysis

in line with ITU-T recommendation P.863

Numeric results

narrowband or super-wideband

MOS-LQO,
level,
attenuation,
SNR (signal to noise ratio),
ASR (active speech ratio),
delay (average, minimum, maximum)

Graphic displays (versus time)

MOS-LQO, delay,
reference signal and degraded signal

16

PESQ

®

, PEAQ

®

and POLQA

®

are registered trademarks of OPTICOM Dipl.-Ing. M. Keyhl GmbH, Germany.