Jitter and interface test (r&s, Upv-k22 option), Generator – Atec Rohde-Schwarz-UPV User Manual
Page 23: Analyzer
![background image](/manuals/318890/23/background.png)
Version 03.00, August 2011
Rohde & Schwarz R&S
®
UPV Audio Analyzer 23
Jitter and interface test (R&S
®
UPV-K22 option)
Digital audio I/O 192 kHz (R&S
®
UPV-B2 option) required.
Generator
Jitter injection
sine
0.1 Hz to 80 kHz
random
12 Hz to 80 kHz
Signals
arbitrary
80 kHz bandwidth,
192 kHz sampling rate,
256 ksample (max.)
Amplitude (peak)
0 to 2.5 UI
Common mode injection
at balanced output
sine
0.1 Hz to 80 kHz
random
12 Hz to 80 kHz
Signals
arbitrary
80 kHz bandwidth,
192 kHz sampling rate,
256 ksample (max.)
Amplitude (peak)
0 V to +10 V
Analyzer
Jitter measurement
Analyzer functions
RMS, RMS selective, peak, frequency,
FFT, waveform
10 Hz to 250 kHz
3 dB bandwidth
> 150 kHz
48 kHz sampling rate
typ. 0.75 UI to 80 kHz,
decreasing to 25 kHz at 2.5 UI
96 kHz sampling rate
typ. 1.25 UI to 80 kHz,
decreasing to 40 kHz at 2.5 UI
Measuring range
192 kHz sampling rate
typ. 1.5 UI to 80 kHz,
decreasing to 50 kHz at 2.5 UI
Level error
±(10 % + 1 ns)
300 Hz to 50 kHz
±10 %
Flatness
50 kHz to 80 kHz
±20 %
Inherent jitter
700 Hz to 80 kHz
< 0.01 UI (peak)
Spurious jitter
700 Hz to 80 kHz
< –35 dBc or < –50 dBUI,
whichever is larger
Common mode test
at balanced input
Analyzer functions
RMS, RMS selective, peak, frequency,
FFT, waveform
Frequency range
10 Hz to 250 kHz
Amplitude range
0 V to 30 V
Input signal
Amplitude (peak-peak)
0 V to 10 V
Sampling rate
30 kHz to 200 kHz