beautypg.com

Jitter and interface test (r&s, Upv-k22 option), Generator – Atec Rohde-Schwarz-UPV User Manual

Page 23: Analyzer

background image

Version 03.00, August 2011

Rohde & Schwarz R&S

®

UPV Audio Analyzer 23

Jitter and interface test (R&S

®

UPV-K22 option)

Digital audio I/O 192 kHz (R&S

®

UPV-B2 option) required.

Generator

Jitter injection

sine

0.1 Hz to 80 kHz

random

12 Hz to 80 kHz

Signals

arbitrary

80 kHz bandwidth,
192 kHz sampling rate,
256 ksample (max.)

Amplitude (peak)

0 to 2.5 UI

Common mode injection

at balanced output
sine

0.1 Hz to 80 kHz

random

12 Hz to 80 kHz

Signals

arbitrary

80 kHz bandwidth,
192 kHz sampling rate,
256 ksample (max.)

Amplitude (peak)

0 V to +10 V

Analyzer

Jitter measurement
Analyzer functions

RMS, RMS selective, peak, frequency,
FFT, waveform

10 Hz to 250 kHz

3 dB bandwidth

> 150 kHz

48 kHz sampling rate

typ. 0.75 UI to 80 kHz,
decreasing to 25 kHz at 2.5 UI

96 kHz sampling rate

typ. 1.25 UI to 80 kHz,
decreasing to 40 kHz at 2.5 UI

Measuring range

192 kHz sampling rate

typ. 1.5 UI to 80 kHz,
decreasing to 50 kHz at 2.5 UI

Level error

±(10 % + 1 ns)

300 Hz to 50 kHz

±10 %

Flatness

50 kHz to 80 kHz

±20 %

Inherent jitter

700 Hz to 80 kHz

< 0.01 UI (peak)

Spurious jitter

700 Hz to 80 kHz

< –35 dBc or < –50 dBUI,
whichever is larger

Common mode test

at balanced input

Analyzer functions

RMS, RMS selective, peak, frequency,
FFT, waveform

Frequency range

10 Hz to 250 kHz

Amplitude range

0 V to 30 V

Input signal
Amplitude (peak-peak)

0 V to 10 V

Sampling rate

30 kHz to 200 kHz