beautypg.com

Analog analyzers, Dual-channel inputs, Dual-channel analog analyzer measurement functions – Atec Rohde-Schwarz-UPV User Manual

Page 5

background image

Version 03.00, August 2011

Rohde & Schwarz R&S

®

UPV Audio Analyzer 5

Analog analyzers

Analog measurements are available with different bandwidths, specifications and measurement functions.

Dual-channel inputs

Analyzer
Bandwidth 22 kHz

DC/10 Hz to 21.76 kHz

1

Bandwidth 40/80 kHz

DC/10 Hz to 40/80 kHz

1

Bandwidth 250 kHz

DC/10 Hz to 250 kHz

1

Level measurements (RMS)
Level error

at 1 kHz

±0.05 dB, typ. ±0.025 dB

20 Hz to 20 kHz

±0.01 dB, typ. 0.003 dB, V

in

< 3 V

2

20

kHz to 50 kHz

±0.03 dB, V

in

< 3 V

2

50 kHz to 100 kHz

±0.1 dB

Frequency response (referenced to 1 kHz)

100 kHz to 250 kHz

±0.3 dB

XLR connectors

2 channels, balanced
(unbalanced measurements possible with the R&S

®

UP-Z1MF XLR/BNC adapter set),

XLR pin 1 floating/grounded selectable,
AC/DC coupling selectable

Voltage range

RMS, sine

0.1 µV to 110 V

Measurement ranges

18 mV to 100 V, in steps of 5 dB
100 kΩ

±1 % shunted by 120 pF (230 pF

for ranges ≥ 6 V), each pin against ground
300 Ω

±0.5 %, P

max

2 W

Input impedance

600 Ω

±0.5 %, P

max

1 W

Crosstalk attenuation

frequency < 22 kHz, 600 Ω

> 120 dB

at 50 Hz

> 90 dB

at 1 kHz

> 86 dB

Common-mode rejection (V

in

< 3 V)

at 20 kHz

> 80 dB

each input channel switchable to any output channel

200 k

Ω balanced

Generator output

input impedance

100 k

Ω unbalanced

Dual-channel analog analyzer measurement functions

RMS, wideband

measurement speed AUTO

±0.05 dB, typ. ±0.025 dB, at 1 kHz, sine

Level error

measurement speed AUTO FAST

±0.1 dB additional error

AUTO FAST/AUTO

min. 200/4000 sample, at least 1 cycle

GEN TRACK

min. 100 sample, at least 1 cycle

Integration time

VALUE

0.1 ms to 100 s

22/40/80 kHz bandwidth
A weighted

< 1 µV, typ. 0.7 µV

CCIR unweighted

< 1.4 µV, typ. 1 µV

80 kHz bandwidth

< 2.8 µV

Noise (input shorted)

250 kHz bandwidth

< 7 µV

Spectrum

post

FFT

DC voltage
Voltage range

0 V to ±110 V

Level error

3

±(1 % of measured value + 0.1 % of
measurement range)

Measurement ranges

100 mV to 100 V, in steps of 10 dB

FFT analysis

see FFT analyzer section

1

DC/AC

coupling.

2

Additionally

± 0.02 dB from 5 kHz to 50 kHz when V

in

≥ 3 V.

3

Not valid for bandwidth 250 kHz.