beautypg.com

Atec Rohde-Schwarz-UPV User Manual

Page 6

background image

Version 03.00, August 2011

6

Rohde & Schwarz R&S

®

UPV Audio Analyzer

Total harmonic distortion (THD)
Fundamental

10 Hz to 110 kHz

Frequency tuning

automatic to input or generator signal or
fixed through entered value

Weighted harmonics

any combination of d

2

to d

9

, up to 250 kHz

harmonics < 50 kHz

±0.5 dB

harmonics < 100 kHz

±0.7 dB

Error limit

harmonics < 250 kHz

±1 dB

fundamental 20 Hz to 10.95 kHz

< –110 dB, typ. –115 dB

Inherent distortion

(bandwidth 22 kHz)

4

5

fundamental 10 Hz to 20 Hz

< –100 dB

fundamental 50 Hz to 20 kHz

< –100 dB, typ. –105 dB

fundamental 20 kHz to 50 kHz

< –90 dB, typ. –95 dB

Inherent distortion

(bandwidth 40/80/250 kHz)

4

5

fundamental 50 kHz to 110 kHz

< –80 dB, typ. –85 dB

Spectrum

bargraph showing signal and distortion
post FFT

THD+N and SINAD
Fundamental

10 Hz to 110 kHz

Frequency tuning

automatic to input or generator signal or
fixed through entered value

Input voltage

typ.

> 100 µV with automatic tuning

Bandwidth

upper and lower frequency limit selectable,
one weighting filter in addition

bandwidth

< 50 kHz

±0.5 dB

< 100 kHz

±0.7 dB

Error limit

< 250 kHz

±1 dB

fundamental meas.

Bandwidth

Inherent distortion

(analyzer bandwidth 22 kHz)

4

to 22 kHz

20 Hz to 22

kHz


typ. –110 dB at 1 kHz, 2.5 V
< –105 dB + 2 µV

6

typ. –108 dB + 1.5 µV

fundamental meas.

bandwidth

to 20kHz

20 Hz to 22 kHz


< –95 dB + 2.5 µV, typ. –100 dB + 1.75 µV

to 20 kHz

20 Hz to 80 kHz

< –90 dB + 5 µV, typ. –95 dB + 3.5 µV

Inherent distortion

(analyzer bandwidth 40/80/250 kHz)

4

to 50 kHz

20 Hz to 250 kHz

< –84 dB + 10 µV, typ. –90 dB + 7 µV

Spectrum

post

FFT

Time domain display (WAVEFORM)
Trigger

rising/falling

Trigger level

–100 V to +100 V

Trace length

max. 480 ksample per channel

Pretrigger

max. 19200 sample

Standard mode

each sample recorded

Compressed mode

peak value of up to 1024 sample recorded
(envelope)

Undersample mode

undersampling factor up to 1024

Frequency
Frequency range

20 Hz to 250 kHz

Frequency error

±10 ppm

Phase
Frequency range

20 Hz to 250 kHz

20 Hz to 22 kHz

±0.4°

22 kHz to 50 kHz

±0.6°

50 kHz to 100 kHz

±1.0°

Phase error

100 kHz to 250 kHz

±1.5°

4

Total inherent distortion of analyzer and generator (with R&S

®

UPV-B1 option), analyzer with dynamic mode precision.

5

Typ. 3 dB less when > 3.5 V; sensitivity reduced by inherent noise when < 0.5 V.

6

At full-scale level of measurement range (< –100 dB + 2 µV with autoranging), < –100 dB for input voltage > 3.5 V.