beautypg.com

Sweep, Display of results – Atec Rohde-Schwarz-UPV User Manual

Page 21

background image

Version 03.00, August 2011

Rohde & Schwarz R&S

®

UPV Audio Analyzer 21

Analog notch filter

for measurements on signals with high
S/N ratio, this filter improves the dynamic
range of the analyzer by up to 30 dB to
140 dB for an analyzer bandwidth of
22/40/80 kHz, or 120 dB for an analyzer
bandwidth of 250 kHz (typical noise floor
of FFT); the filter is also used for
measuring THD, THD+N and MOD DIST
with dynamic mode precision

RMS (wideband)
RMS (selective)
quasi-peak

Characteristics available

in dual channel analog analyzer

with measurement functions

FFT analysis

Frequency range

center frequency (f

c

)

10 Hz to 110 kHz
automatic to input signal
coupled to generator

Frequency tuning

fixed through entered value

Stopband f

c

±

0.5 %

typ. >

30 dB

at 0.77 × f

c

and 1.3 × f

c

typ. –3 dB

Passband

outside 0.5 × f

c

to 2 × f

c

typ. +0 dB/–1 dB

Sweep

Generator sweep
Parameters

frequency, level, with bursts also interval
and duration, one- or two-dimensional

Sweep

linear, logarithmic, single, continuous
automatic after end of measurement

Stepping

time delay (fixed or loaded table)

Sweep speed

GEN TRACK

1.0 s

AUTO FAST

1.2 s

Two-channel RMS measurement 20 Hz to
20 kHz, 30-point generator sweep
logarithmic (frequency measurement
switched off, Low Dist OFF)

AUTO 2.0

s

Display of results

Units
Level (analog)

V, dBu, dBV, W, dBm, difference (

Δ),

deviation (

Δ%) and ratio (without

dimension, %, dBr) to reference value

Level (digital)

FS, %FS, dBFS, LSBs,
deviation (

Δ%) or ratio (dBr)

to reference value

Distortion

% or dB, referenced to signal amplitude,
THD and THD+N in all available level units
(absolute or relative to selectable
reference value)

Frequency

Hz, difference (

Δ), deviation (Δ%) and ratio

(as quotient f/f

ref

, 1/3 octave,

octave or decade) to reference value
(entered or stored, current generator
frequency)

Phase

°, rad, difference (Δ) to reference value
(entered or stored)

Reference value (level)

fixed value (entered or stored)