beautypg.com

Signals – Atec Rohde-Schwarz-UPV User Manual

Page 18

background image

Version 03.00, August 2011

18

Rohde & Schwarz R&S

®

UPV Audio Analyzer

Signals

All signals 24 bit, full scale.

General characteristics

for sine, stereo sine, DFD and MOD DIST

distribution

Gaussian, triangular, rectangular

Dither

level

0.5 LSB to 1 FS

internal clock

±10 ppm

Frequency error

relative to clock rate

±1 ppm

DC offset

0 to ±1 FS adjustable

Sine
Frequency range

0.1 Hz to 49.9 % of sampling rate

Inherent distortion (THD)

14

< –155 dB

Sweep parameters

frequency, level

Stereo sine
Frequency range

0.1 Hz to 47.9 % of sampling rate

Frequency

adjustable for each channel

Phase

same frequency in both channels

–360° to +360°

Level

adjustable for each channel
or channel ratio 2/1

Sweep parameters

frequency and level of channel 1

MOD DIST

for measuring the modulation distortion

30 Hz to UF/8

Frequency range

lower frequency (LF)
upper frequency (UF)

8 × LF to 47.9 % of sampling rate

Level ratio (LF:UF)

selectable from 10:1 to 1:1

Inherent distortion

14

level LF:UF = 4:1
with 1 LSB triangular dither

< –142 dB

Sweep parameters

upper frequency, level

DFD

for measuring the difference frequency distortion

difference frequency

80 Hz to 2 kHz

Frequency range

mean frequency

200 Hz

to 20.90 kHz

Inherent distortion

14

DFD

d

2

, DFD d

3

with 1 LSB triangular dither

< –155 dB

Sweep parameters

mean frequency, level

Sine burst, sine

2

burst

Burst time

1 sample up to 60 s, 1-sample-resolution

Interval

burst time up to 3600 s,
1-sample-resolution

Low level

0 to burst level, absolute or referenced to
burst level (0 for sine

2

burst)

Sweep parameters

burst frequency, level, time, interval

Noise
Distribution

Gaussian, triangular, rectangular

Arbitrary waveform
Memory depth

max. 256 ksample

Clock rate

sampling rate of generator

File format

*.arb

14

Total inherent distortion of analyzer and generator.