Definitions and Conditions

Specifications represent warranted performance of a calibrated instrument that

has been stored for a minimum of 2 hours within the operating temperature range

of 0 to 55 °C, unless otherwise stated, and after a 45 minute warm-up period . The

specifications include measurement uncertainty . Data represented in this document

are specifications unless otherwise noted .

Typical (typ) describes additional product performance information that is not

covered by the product warranty . It is performance beyond specifications that 80

percent of the units exhibit with a 90 percent confidence level at room temperature

(approximately 25 °C) . Typical performance does not include measurement

uncertainty .

Nominal (nom) values indicate the expected mean or average performance, or an

attribute whose performance is by design, such as the 50 ohm connector . This data

is not warranted and is measured at room temperature (approximately 25 °C) .

Measured (meas) describes an attribute measured during the design phase for

purposes of communicating expected performance, such as amplitude drift vs . time .

This data is not warranted and is measured at room temperature (approximately

25 °C) .