beautypg.com

E-series power sensor specifications, E-series cw power sensor specifications – Atec Agilent-E9300A User Manual

Page 9

background image

9

E-series power sensor
specifications

The E-series of power sensors have
their calibration factors stored in
EEPROM and operate over a wide
dynamic range. They are designed
for use with the EPM series of power
meters and two classes of sensors
are available:

• CW power sensors (E4412A and

E4413A).

• Average power sensors (E9300

sensors).

Power linearity

Table 7
Power

Temperature (25ºC ±5ºC)

Temperature (0º to 55ºC)

100 pW to 10 mW

±3 %

±7 %

(-70 dBm to +10 dBm)
10 mW to 100 mW

±4.5%

±10%

(+10 dBm to +20 dBm)

E-series CW power sensor specifications

Wide dynamic range CW sensors: 100 pW to 100 mW (-70 dBm to +20 dBm)

Table 6
Model

Frequency range

Maximum SWR

Maximum power

Connector type

E4412A

10 MHz – 18 GHz

*10 MHz to 30 MHz: 1.12

200 mW (+23 dBm) Type-N (m)

30 MHz to 2 Ghz:1.15
2 GHz to 6 Ghz:1.17
6 GHz to 11 Ghz:1.2
11 GHz to 18 Ghz:1.27

E4413A

50 MHz – 26.5 GHz

50 MHz to 100 MHz: 1.21 200 mW (+23 dBm) APC-3.5mm (m)
100 MHz to 8 Ghz:1.19
8 GHz to 18 Ghz:1.21
18 GHz to 26.5 Ghz:1.26

* Applies to sensors with serial prefix US 3848 or greater

Figure 1 : Relative mode power measurement
linearity with EPM series power meter/E-series
CW power sensor at 25°C ± 5°C (typical)