3
Supplemental characteristics
Power reference
Frequency: 50 MHz nominal.
SWR: 1.05 maximum.
Connector: Type N (f), 50
Ω
.
Measurement speed
Three measurement speed modes,
over the GPIB, are available as
shown, along with the typical maxi-
mum measurement speed for each
mode.
With the E4418B power meter
Normal: 20 readings/second
x2: 40 readings/second
Fast: 200 readings/second
With the E4419B, the measurement
speed is reduced, for example, with
both channels in FAST mode, the
typical maximum measurement
speed is 100 readings/second.
Fast mode is for E-series sensors
only.
Maximum measurement speed is
obtained using binary output
and in free run trigger mode.
Zero drift of sensors: Sensor depen-
dent, refer to table 2. For E9300
sensors, refer to table 12 for
complete data.
Measurement noise: Sensor depen-
dent, refer to table 2 and table 3. For
E9300 sensors, refer to table 12.
Effects of averaging on noise: Averaging
over 1 to 1024 readings is available
for reducing noise. Table 2 provides
the measurement noise for a particu-
lar sensor with the number of
averages set at 16 (for normal mode)
and 32 (for x2 mode). Use the noise
multiplier, for the appropriate mode
(normal or x2) and number of
averages, to determine the total mea-
surement noise value.
Example:
8481D power sensor, normal mode,
number of averages = 4.
Measurement noise calculation:
(< 45 pW x 2.75) = < 121 pW.
Table 2
Model Zero
drift
1
Measurement noise2
8481A, 8482A, 8483A, 8485A,
< ±10 nW
< 110 nW
8487A, R8486A, Q8486A
8481B, 8482B
< ±10 µW
< 110 µW
8481D, 8485D, 8487D
< ±4 pW
< 45 pW
8481H, 8482H
< ±1µW
< 10 µW
R8486D, Q8486D
< ±6 pW
< 65 pW
V8486A, W8486A
< ±40 nW
< 450 nW
E4412A, E4413A
< ±15 pW
< 70 pW
E9300A, E9301A, E9304A3
< ±150 pW
< 700 pW
E9300B, E9301B3
< ±150 nW
< 700 nW
E9300H, E9301H3
< ±1.5 nW
< 7 nW
1. Within 1 hour after zero set, at a constant temperature, after a 24-hour warm-up of the power
meter.
2. The number of averages at 16 for normal mode and 32 for x2 mode, at a constant temperature,
measured over a one minute interval and two standard deviations. For E-series sensors, the
measurement noise is measured within the low range. Refer to the relevant sensor manual for
further information.
3. Specification applies to the low power path, 15% to 75% relative humidity.
Table 3
Number of averages
1
2
4
8
16
32
64
128
256
512
1024
Noise
multiplier
5.5 3.89 2.75 1.94 1
0.85 0.61 0.49 0.34 0.24 0.17
(normal mode)
Noise
multiplier
6.5 4.6 3.25 2.3 1.63 1
0.72 0.57 0.41 0.29 0.2
(x2 mode)