Atec Agilent-E9300A User Manual
Page 19
19
8480 series power sensors
(with EPM series power
meters)
The 8480 series power sensors are
designed for use with the 435B,
436A, 437B, 438A, 70100A, E1416A
and now the E4418B and E4419B
power meters. These thermocouple
and diode power sensors provide
extraordinary accuracy, stability,
and SWR over a wide range of
frequencies (100 kHz to 110 GHz)
and power levels -70 dBm to
+44 dBm).
Table 14. Root sum of squares (rss) uncertain-
ty on the calibration factor data printed on the
power sensor
Freq 8482A
8482B
8482H
8483A
(MHz)
0.1 1.3 2.8 1.6 1.5
0.3 1.2 2.8 1.6 1.4
1 1.2
2.8
1.6
1.4
3 1.2
2.8
1.6
1.4
10 1.3 2.8 1.6
1.6
30 1.4 2.8 1.7
1.6
50
0 (ref)
2.7
0 (ref) 0 (ref)
100 1.6 3.3 1.9 2
300 1.6 3.3 1.9 2
1000 1.4 3.3 1.7 2
2000 1.4 3.3 1.7 2.1
4000 1.5 3.1 1.8 -
Freq
(GHz)
8481A 8481B 8481H 8481D 8485A 8385D 8487A 8487D
1
1.6 3 1.9 1.9 1.6 1.8 1.6 2
2
1.4 3.1 1.7 1.8 1.6 1.8 1.6 2
4
1.5 3.1 1.8 1.8 1.7 1.8 1.6 2
6
1.5 3.1 1.8 1.8 1.8 2.1 1.7 2.3
8
1.7
3.2
2 2 1.9 2.2
1.8 2.3
10
1.9 3.3 2.2 2.2 2
2.1 1.8 2.3
12
2.1 4.1 2.4 2.8 2
2.2 1.9 2.3
14
2.6 4.1 2.8 3.2 2.2 2.2 2.1 2.8
16
2.9 4.2 3 3.4 2.3 2.5 2.2 2.8
18
3.2 4.3 3.1 3.7 2
2.6 2.3 2.8
22 - - - - 2.1
2 1.8
2.8
26.5 - - - - 2.1
2.3
2.1
2.8
Freq (GHz) R8486A Q8486A R8486D Q8486D 8487A
8487D
26.5 2.2
- 3 - 2.1
2.8
28 2.4
- 3.2
- 2.3
3
30 2.5
- 3 - 2.1
3
33
2.1 2.8 3 4.2 2.3 3
34.5 2.1 2.8 3 4.2 2.1 3
37
2.2 2.8 3 4.2 2.3 3
40
2.2 2.9 3 4.2 2.6 3
42
- 3.9 - 4.9 3.2 2.9
44
- 3.9 - 5.1 3.6 2.9
46
- 3.9 - 5.5 4.1 3.1
48
- 4.9 - 5.8 4.5 4.5
50
- 5.3 - 6.2 5
4.5