11

E-series E9300
average power sensor
specifications

The E-series E9300 wide dynamic
range, average power sensors are
designed for use with the EPM fami-
ly of power meters. These specifica-
tions are valid ONLY after proper
calibration of the power meter and
apply for CW signals unless other-
wise stated.

Specifications apply over the tem-
perature range 0°C to 55°C unless
otherwise stated, and specifications
quoted over the temperature range
25°C ±10°C, conform to the stan-
dard environmental test conditions
as defined in TIA/EIA/ IS-97-A and
TIA/EIA/IS-98-A [1].

The E-series E9300 power sensors
have two independent measurement
paths (high and low power paths):

High power path: -10 to +20 dBm (“A” suffix sensors), +20 to +44 dBm (“B” suffix sensors) and 0 to +30 dBm
(“H” suffix sensors).

Low power path: -60 to -10 dBm (“A” suffix sensors), -30 to +20 dBm (“B” suffix sensors) and -50 to 0 dBm
(“H” suffix sensors).

Some specifications are detailed for an individual measurement path.

Table 9
Wide dynamic range (-60 to +20 dBm) sensors
Model Frequency range

Maximum SWR

Maximum SWR

Maximum power

Connector

(25°C ± 10°C)

(0 – 55°C)

type

E9300B 10 MHz – 18.0 GHz 10 MHz to 30 GHz: 1.15

10 MHz to 30 MHz: 1.21

+25 dBm average;

Type-N (m)

30 MHz to 2 GHz: 1.13

30 MHz to 2 GHz: 1.15

+33 dBm peak

2 GHz to 14 GHz: 1.19

2 GHz to 14 GHz: 1.20

(< 10 µsec)

14 GHz to 16 GHz: 1.22

14 GHz to 16 GHz: 1.23

16 GHz to 18 GHz: 1.26

16 GHz to 18 GHz: 1.27

E9301A 10 MHz – 6.0 GHz

10 MHz to 30 GHz: 1.15

10 MHz to 30 MHz: 1.21

+25 dBm average;

Type-N (m)

30 MHz to 2 GHz: 1.13

30 MHz to 2 GHz: 1.15

+33 dBm peak

2 GHz to 14 GHz: 1.19

2 GHz to 14 GHz: 1.20

(< 10 µsec)

E9304A 9 MHz – 6.0 GHz

9 kHz to 2 GHz: 1.13

9 kHz to 2 GHz: 1.15

+25 dBm average;

Type-N (m)

2 GHz to 6 GHz: 1.19

2 GHz to 6 GHz: 1.20

+33 dBm peak
(< 10 µsec)

Wide dynamic range (-30 to +44 dBm) sensors
Model Frequency range

Maximum SWR

Maximum SWR

Maximum power

Connector

(25°C ± 10°C)

(0 – 55°C)

type

E9300B 10 MHz – 18.0 GHz 10 MHz to 8 GHz: 1.12

10 MHz to 8 GHz: 1.14

0 – 35°C: 30 W avg

Type-N (m)

8 to 12.4 GHz: 1.17

8 to 12.4 GHz: 1.18

35 – 55°C: 25 W avg

12.4 to 18 GHz: 1.24

12.4 to 18 GHz: 1.25

< 6 GHz: 500 W pk
> 6 GHz: 125 W pk
500 W-mS per pulse

E9301B 10 MHz – 6.0 GHz

10 MHz to 6 GHz: 1.12

10 MHz to 6 GHz: 1.14

0 – 35°C: 30 W avg

Type-N (m)

35 – 55°C: 25 W avg
< 6 GHz: 500 W pk
500 W-mS per pulse

Wide dynamic range (-50 to +30 dBm) sensors
Model Frequency range

Maximum SWR

Maximum SWR

Maximum power

Connector

(25°C ± 10°C)

(0 – 55°C)

type

E9300H 10 MHz – 18.0 GHz 10 MHz to 8 GHz: 1.15

10 MHz to 8 GHz: 1.17

3.16 W avg

Type-N (m)

8 to 12.4 GHz: 1.25

8 to 12.4 GHz: 1.26

100 W pk

12.4 to 18 GHz: 1.28

12.4 to 18 GHz: 1.29

100 W-µS per pulse

E9301H 10 MHz – 6.0 GHz

10 MHz to 6 GHz: 1.15

10 MHz to 6 GHz: 1.17

3.16 W avg

Type-N (m)

100 W pk
100 W-µS per pulse