15

E-series E9300 average power sensors

Typical E9300B/01B power linearity at 25°C, after zero and calibration, with associated
measurement uncertainty

Power range

-6 to 0 dBm

0 to +10 dBm

+10 to +20 dBm

+20 to +26 dBm

Measurement uncertainty

± 0.65% ± 0.55%

± 0.45% ± 0.31%

Typical E9300H/01H power linearity at 25°C, after zero and calibration, with associated
measurement uncertainty

Power range

-26 to -20 dBm -20 to –10 dBm -10 to 0 dBm 0 to +10 dBm +10 to +20 dBm

+20 to +26 dBm

Measurement uncertainty

± 0.9% ± 0.8% ± 0.65%

± 0.55%

± 0.45% ± 0.31%