beautypg.com

E-series e9300 average power sensors – Atec Agilent-E9300A User Manual

Page 14

background image

14

E-series E9300 average power sensors

Table 10. Power linearity (after zero and cal at ambient environmental conditions)
Sensor Power

Linearity

Linearity

(25°C ± 10°C)

(0 – 55°C)

E9300A, E9301A, E9304A

-60 to –10 dBm

±3.0%

±3.5%

-10 to 0 dBm

±2.5%

±3.0%

0 to +20 dBm

±2.0%

±2.5%

E9300B, E9301B

-30 to +20 dBm

±3.5%

±4.0%

+20 to +30 dBm

±3.0%

±3.5%

+30 to +44 dBm

±2.5%

±3.0%

E9300H, E9301H

-50 to 0 dBm

±4.0%

±5.0%

0 to +10 dBm

±3.5%

±4.0%

+10 to +30 dBm

±3.0%

±3.5%

Typical E9300A/01A/04A Power Linearity at 25°C, after zero and calibration, with associated
measurement uncertainty

Power range

-30 to -20 dBm

-20 to -10 dBm

-10 to 0 dBm

0 to +10 dBm

+10 to +20 dBm

Measurement uncertainty

±0.9% ±0.8% ±0.65%

±0.55%

±0.45%