8480 series sensors (with epm series power meters) – Atec Agilent-E9300A User Manual
Page 21
21
8480 series sensors (with EPM series power meters)
High sensitivity sensors, 100 pW to 10 µW (–70 dBm to –20 dBm)
Model
Frequency range
Maximum SWR
Power linearity1
Maximum power
Connector type
Weight
8481D3
10 MHz to 18 GHz
10 MHz to 30 MHz: 1.40
- 30 dBm to - 20 dBm: (±1%)
100 mW avg, 100 m W pk
Type - N (m)
Net: 0.16 kg (0.37 lb)
30 MHz to 3 GHz: 1.15
Shipping: 0.9 kg (2.0 lb)
4 GHz to 10 GHz: 1.20
10 GHz to 15 GHz: 1.30
15 GHz to 18 GHz: 1.50
8485D3
50 MHz to 26.5 GHz
0.05 GHz to 0.1 GHz: 1.19
-30 dBm to -20 dBm: (±2%)
100 mW avg, 100 m W pk
APC - 3.5mm (m)
Net: 0.2 kg (.38 lb)
0.1 GHz to 4 GHz: 1.15
Shipping: 0.5 kg (1.0 lb)
4 GHz to 12 GHz: 1.19
12 GHz to 18 GHz: 1.25
18 GHz to 26.5 GHz: 1.29
Option
50 MHz to 33 GHz
26.5 GHz to 33 GHz: 1.35
-30 dBm to -20 dBm: (±2%)
100 mW avg, 100 m W pk
APC - 3.5mm (m)
Net: 0.2 kg (.38 lb)
8485D-033
Shipping: 0.5 kg (1.0 lb)
8487D3
50 kHz to 50 GHz
0.05 GHz to 0.1 GHz: 1.19
-30 dBm to -20 dBm: (±2%)
100 mW avg, 100 m W pk
2.4 mm (m)
Net: 0.2 kg (0.38 lb)
0.1 GHz to 4 GHz: 1.15
30 W.µs per pulse
Shipping: 0.5 kg (1.0 lb)
2 GHz to 12.4 GHz: 1.20
12.4 GHz to 18 GHz: 1.29
18 GHz to 34 GHz: 1.37
34 GHz to 40 GHz: 1.61
40 GHz to 50 GHz: 1.89
R8486D3
26.5 GHz to 40 GHz
26.5 GHz to 40 GHz: 1.40
-30 dBm to -25 dBm: (±3%)
100 mW avg, or pk
Waveguide flange
Net: 0.26 kg (0.53 lb)
-25 dBm to -20 dBm: (±5%)
40 V dc max
UG-599/U
Shipping: .66 kg (1.3 lb)
Q8486D3
33 GHz to 50 GHz
33 GHz to 50 GHz: 1.40
-30 dBm to 25 dBm: (±3%)
100 mW avg, or pk
Waveguide flange
Net: 0.26 kg (0.53 lb)
-25 dBm to -20 dBm: (±5%)
40 V dc max
UG-383/U
Shipping: 0.66 kg (1.3 lb)
1. Negligible deviation except for those power ranges noted.
2. For pulses greater than 30 W, the maximum average power (Pa) is limited by the energy per pulse (E) in W.µs according to Pa = 30-0.02 E.
3. Includes 11708A 30 dB attenuator for calibrating against 0 dBm, 50 MHz power reference. The 11708A is factory set to 30 dB ±0.05 dB at 50 MHz,
traceable to NIST. SWR < 1.05 at 50 MHz.