EPM series E4418B and E4419B power meters

Specifications

Specifications describe the instru-
ment’s warranted performance and
apply after a 30 minute warm-up.
These specifications are valid over
its operating/environmental range
unless otherwise stated and after
performing a zero and calibration
procedure.

Supplemental
characteristics

Supplemental characteristics (shown
in italics) are intended to provide
additional information, useful in
applying the instrument by giving
typical (expected), but not warrant-
ed performance parameters. These
characteristics are shown in italics
or labeled as “typical”, “nominal” or
“approximate”.

For information on measurement
uncertainty calculations, refer to
Application Note 64-1C, “Funda-
mentals of RF and Microwave
Power Measurements
”, literature
number 5965-6380E.

Frequency range: 9 kHz to 110
GHz, sensor dependent

Power range: –70 dBm to +44 dBm
(100 pW to 25 W), sensor dependent.

Power sensors: Compatible with all
8480 series and E-series sensors

Single sensor dynamic range:

90 dB maximum (E-series sensors),
50 dB maximum ( 8480 series
sensors).

Display units:

Absolute: Watts or dBm.
Relative: Percent or dB.

Display resolution: Selectable
resolution of 1.0, 0.1, 0.01, and 0.001
dB in Log mode, or 1 to 4 digits in
linear mode.

Default resolution: 0.01 dB in log
mode, 3 digits in linear mode.

Accuracy

Instrumentation

Absolute: ±0.02 dB (log) or ±0.5%
(linear). Please add the correspond-
ing power sensor linearity percent-
age from tables 7 and 10 (for the
E-series sensors) and table 15 (for
the 8480 series sensors).

Relative: ±0.04 dB (Log) or ±1.0%
(linear). Please add the correspond-
ing power sensor linearity percent-
age from table 16 (for the 8480
series sensors).

Zero set (digital settability of zero):
sensor dependent (refer to
table 1). For E-series sensors, this
specification applies to a ZERO
performed when the sensor input is
not connected to the POWER REF.

Power reference

Power output: 1.00 mW (0.0 dBm).
Factory set to ±0.4% traceable to the
National Physical Laboratory (NPL),
UK

1

.

Accuracy: For two years:
±0.5% (23 ± 3°C)
±0.6% (25 ± 10°C)
±0.9% (0 to 55°C)

Frequency: 50 MHz nominal

SWR: 1.06 maximum
(1.08 maximum for Option 003)

Connector type: Type N (f), 50

Table 1
Model ZERO

SET

8481A, 8482A, 8483A, 8485A,

±50 nW

8487A, R8486A, Q8486A,

8481B, 8482B

±50 µW

8481D, 8485D, 8487D

±20 pW

8481H, 8482H

±5 µW

R8486D, Q8486D

±30 pW

V8486A, W8486A

±200 nW

E4412A, E4413A

±50 pW

E9300A, E9301A,E9304A 2

±500 pW

E9300B, E9301B 2

±500 nW

E9300H, E9301H 2

±5 nW

1. National metrology institutes of member states

of the Metre Convention, such as the National
Institute of Standards and Technology in the
USA, are signatories to the ComitÈ International
des Poids et Mesures Mutual Recognition
Arrangement. Further information is available
from the Bureau International des Poids et
Mesures, at http://www.bipm.fr/

2. Specification applies to the low power path

15% to 75% relative humidity.