Continued), Settling time (continued) – Atec Agilent-EPM Series User Manual

Page 7

background image

7

+40 dBm

+32 dBm

+26 dBm

+20 dBm

+10 dBm

0 dBm

–10 dBm

–20 dBm

+20 dBm

+12 dBm

+6 dBm

0 dBm

–10 dBm

–20 dBm

–30 dBm

–40 dBm

Typical
settling
times

x2

speed

Normal

speed

40 ms

70 ms

120 ms

210 ms

210 ms

400 ms

400 ms

1 s

40 ms

70 ms

70 ms

120 ms

400 ms

1 s

3.4 s

6.5 s

6.8 s

13 s

Sensor
dynamic
range

Maximum dBm

Minimum dBm

Low power
path

High power
path

With E-Series E9300B/01B/00H/01H sensor

+10 dBm

+2 dBm

–4 dBm

–10 dBm

–20 dBm

–30 dBm

–40 dBm

–50 dBm

Typical
settling
times

x2

speed

Normal

speed

40 ms

70 ms

120 ms

210 ms

210 ms

400 ms

400 ms

1 s

40 ms

70 ms

70 ms

120 ms

400 ms

1 s

3.4 s

6.5 s

6.8 s

13 s

Sensor
dynamic
range

Maximum dBm

Minimum dBm

Low power
path

High power
path

With E-Series E9300A/01A/04A sensors

20 dB

10 dB

10 dB

10 dB

Typical
settling
times

x2

speed

Normal

speed

150 ms

150 ms

180 ms

200 ms

350 ms

500 ms

3.5 s

6.6 s

Sensor
dynamic
range

Maximum dBm

Minimum dBm

With 8480 Series sensors

–40 dBm

–50 dBm

–60 dBm

Typical
settling
times

x2

speed

Normal

speed

40 ms

70 ms

70 ms

120 ms

210 ms

400 ms

3.4 s

6.5 s

Sensor
dynamic
range

Maximum dBm

Minimum dBm

With E-Series E4412/13A sensors

20 dB

10 dB

10 dB

5 dB

Typical
settling
times

x2

speed

Normal

speed

150 ms

150 ms

400 ms

1 s

3.6 s

6.6 s

6.6 s

13.5 s

Sensor
dynamic
range

Maximum dBm

Minimum dBm

With N8480 Series sensors

10 dB

180 ms

200 ms

N1913A/14A EPM Series Power Meters Performance Characteristics

(continued)

Settling time (continued)

Auto filter, 10 dB decreasing power step for normal and X2 modes (not across
the range switch points for E-Series and N8480 Series sensors).

This manual is related to the following products: