E-series power sensor specifications, E-series cw power sensor specifications, Continued) – Atec Agilent-EPM Series User Manual
Page 11
11
E-Series Power Sensor Specifications
The E-Series of power sensors have their calibration factors stored in EEPROM
and operate over a wide dynamic range. They are designed for use with the EPM
Series of power meters and two classes of sensors are available:
• CW power sensors (E4412A and E4413A)
• Average power sensors (E9300 sensors)
E-Series CW Power Sensor Specifications
Widest dynamic range: 100 pW to 100 mW (–70 dBm to +20 dBm)
Table 4. E4410 Series max SWR specification
Model
Maximum SWR
Maximum SWR
Maximum power
Connector type
E 4412A
10 MHz to 18 GHz
*10 MHz to <30 MHz: 1.22
30 MHz to <2 GHz: 1.15
2 GHz to <6 GHz: 1.17
6 GHz to <11 GHz:1.2
11 GHz to <18 GHz: 1.27
200 mW (+23 dBm)
Type-N (m)
E4413A
50 MHz to 26.5 GHz
50 MHz to <100 MHz: 1.21
100 MHz to <8 GHz: 1.19
8 GHz to <18 GHz: 1.21
18 GHz to 26.5 GHz: 1.26
200 mW (+23 dBm)
APC-3.5 mm (m)
* Applies to sensors with serial prefix US 3848 or greater
N1913A/14A EPM Series Power Meters Ordering Information
(continued)
Documentation
N191xA-0B0
Delete hard copy English language User’s Guide
N191xA-0BF
Hard copy English language Programming Guide
N191xA-0B1
Hard copy English language User's Guide and Installation Guide
N191xA-0BK
Hard copy English language User's Guide and Programming Guide
N191xA-ABJ
Hard copy Japanese localization User’s Guide and Programming Guide
Options
(continued)