E-Series Power Sensor Specifications

The E-Series of power sensors have their calibration factors stored in EEPROM
and operate over a wide dynamic range. They are designed for use with the EPM
Series of power meters and two classes of sensors are available:

• CW power sensors (E4412A and E4413A)

• Average power sensors (E9300 sensors)

E-Series CW Power Sensor Specifications

Widest dynamic range: 100 pW to 100 mW (–70 dBm to +20 dBm)

Table 4. E4410 Series max SWR specification

Model

Maximum SWR

Maximum SWR

Maximum power

Connector type

E 4412A

10 MHz to 18 GHz

*10 MHz to <30 MHz: 1.22
30 MHz to <2 GHz: 1.15
2 GHz to <6 GHz: 1.17
6 GHz to <11 GHz:1.2
11 GHz to <18 GHz: 1.27

200 mW (+23 dBm)

Type-N (m)

E4413A

50 MHz to 26.5 GHz

50 MHz to <100 MHz: 1.21
100 MHz to <8 GHz: 1.19
8 GHz to <18 GHz: 1.21
18 GHz to 26.5 GHz: 1.26

200 mW (+23 dBm)

APC-3.5 mm (m)

* Applies to sensors with serial prefix US 3848 or greater

N1913A/14A EPM Series Power Meters Ordering Information

(continued)

Documentation

N191xA-0B0

Delete hard copy English language User’s Guide

N191xA-0BF

Hard copy English language Programming Guide

N191xA-0B1

Hard copy English language User's Guide and Installation Guide

N191xA-0BK

Hard copy English language User's Guide and Programming Guide

N191xA-ABJ

Hard copy Japanese localization User’s Guide and Programming Guide

Options

(continued)