E-series e9300 average power sensor specifications, Continued), Power linearity – Atec Agilent-EPM Series User Manual
Page 16: Power range measurement uncertainty
16
E-Series E9300 Average Power Sensor Specifications
(continued)
Power linearity*
Table 9. E9300 Series power linearity (after zero and cal at ambient environmental conditions) sensor
Sensor
Power
Linearity (25 °C ±10 °C)
Linearity (0 °C to 55 °C)
E9300A, E9301A, E9304A
–60 to –10 dBm
–10 to 0 dBm
0 to +20 dBm
±3.0%
±2.5%
±2.0%
±3.5%
±3.0%
±2.5%
E9300B, E9301B
–30 to +20 dBm
+20 to +30 dBm
+30 to +44 dBm
±3.5%
±3.0%
±2.5%
±4.0%
±3.5%
±3.0%
E9300H, E9301H
–50 to 0 dBm
0 to +10 dBm
+10 to +30 dBm
±4.0%
±3.5%
±3.0%
±5.0%
±4.0%
±3.5%
* After zero and calibration at ambient environmental conditions
Typical E9300A/01A/04A power linearity at 25 °C, after zero and
calibration, with associated measurement uncertainty
Power range
Measurement uncertainty
–30 to –20 dBm
±0.9%
–20 to –10 dBm
±0.8%
–10 to 0 dBm
±0.65%
0 to +10 dBm
±0.55%
+10 to +20 dBm
±0.45%
0.5
0.4
0.3
0.2
0.1
0
–0.1
–0.2
–0.3
–0.4
–0.5
% Error
18
14
10
6
2
–2
–8
–10
–14
–18
–22
–26
–30
20
Power (dBm)
1
0.8
0.6
0.4
0.2
0
–0.2
–0.4
–0.6
–0.8
–1
% Error
–10
–5
0
5
10
15
20
25
30
Power (dBm)
1
0.8
0.6
0.4
0.2
0
–0.2
–0.4
–0.6
–0.8
–1
% Error
–10
–5
0
5
10
15
20
25
30
Power (dBm)
Typical E9300B/01B power linearity at 25 °C, after zero and
calibration, with associated measurement uncertainty
Power range
Measurement uncertainty
–6 to 0 dBm
± 0.65%
0 to +10 dBm
± 0.55%
+10 to +20 dBm
± 0.45%
+20 to +26 dBm
± 0.31%
Typical E9300H/01H power linearity at 25 °C, after zero and
calibration, with associated measurement uncertainty
Power range
Measurement uncertainty
–26 to –20 dBm
± 0.9%
–20 to –10 dBm
± 0.8%
–10 to 0 dBm
± 0.65%
0 to +10 dBm
± 0.55%
+10 to +20 dBm
± 0.45%
+20 to +26 dBm
± 0.31%