N1913A/14A EPM Series Power Meters Performance Characteristics

(continued)

Table 3. Settling time

Number of averages

1

2

4

8

16

32

64

128

256

512

1024

Settling time with E-Series sensors (s)

Normal mode
x2 mode

0.08
0.07

0.13
0.09

0.24
0.15

0.45
0.24

1.1
0.45

1.9
1.1

3.5
1.9

6.7
3.6

14
6.7

27
14

57
27

Settling time with N8480 Series sensors (s)

Normal mode
x2 mode

0.15
0.15

0.2
0.18

0.3
0.22

0.5
0.35

1.1
0.55

1.9
1.1

3.4
1.9

6.6
3.5

13
6.9

27
14.5

57
33

Settling time with 8480 Series sensors (s)

Normal mode
x2 mode

0.15
0.15

0.2
0.18

0.3
0.22

0.5
0.35

1.1
0.55

1.9
1.1

3.4
1.9

6.6
3.5

13
6.9

27
14.5

57
33

E-Series sensors In FAST mode (using free run trigger), within the range –50 dBm to +17 dBm, for a 10 dB decreasing
power step, the settling time is:
N1913A: 10 ms

2

N1914A: 20 ms

2

1. Settling time: 0 to 99% settled readings over the GPIB.
2. When a power step crosses through the sensor’s auto-range switch point, add 25 ms. Refer to the relevant sensor manual for switch point information.

Settling time

1

Manual filter, 10-dB decreasing power step for normal and x2 modes (not across
range switch points for E-Series and N8480 Series sensors).