U2000 series usb power sensor specifications, Continued) – Atec Agilent-EPM Series User Manual
Page 26
26
U2000 Series USB Power Sensor Specifications
(continued)
Maximum SWR
Table 19. U2000 Series USB sensors maximum SWR
Model
Frequency range
Maximum SWR
(25 °C ± 10 °C)
Maximum SWR
(0 °C to 55 °C)
U2000A
10 MHz to 30 MHz
30 MHz to 2 GHz
2 GHz to 14 GHz
14 GHz to 16 GHz
16 GHz to 18 GHz
1.15
1.13
1.19
1.22
1.26
1.21
1.15
1.20
1.23
1.27
U2001A
10 MHz to 30 MHz
30 MHz to 2 GHz
2 GHz to 6 GHz
1.15
1.13
1.19
1.21
1.15
1.20
U2002A
50 MHz to 2 GHz
2 GHz to 14 GHz
14 GHz to 16 GHz
16 GHz to 18 GHz
18 GHz to 24 GHz
1.13
1.19
1.22
1.26
1.30
1.15
1.20
1.23
1.27
1.30
U2004A
9 kHz to 2 GHz
2 GHz to 6 GHz
1.13
1.19
1.15
1.20
U2000B
10 MHz to 2 GHz
2 GHz to 12.4 GHz
12.4 GHz to 18 GHz
1.12
1.17
1.24
1.14
1.18
1.25
U2001B
10 MHz to 2 GHz
2 GHz to 6 GHz
1.12
1.17
1.14
1.18
U2000H
10 MHz to 8 GHz
8 GHz to 12.4 GHz
12.4 GHz to 18 GHz
1.15
1.25
1.28
1.17
1.26
1.29
U2001H
10 MHz to 6 GHz
1.15
1.17
U2002H
50 MHz to 8 GHz
8 GHz to 12.4 GHz
12.4 GHz to 18 GHz
18 GHz to 24 GHz
1.15
1.25
1.28
1.30
1.17
1.26
1.29
1.31
For more detailed specifications, refer
to Agilent U2000 Series USB Power
Sensors, data sheet (5989-6278EN).