Calibration factor uncertainties – Atec Agilent-EPM Series User Manual
Page 20
20
848xD Series Diode and 8483A Thermocouple Power Sensor Specifications
Calibration factor uncertainties
These thermocouple and diode power sensors provide extraordinary accuracy,
stability, and SWR over a wide range of frequencies (100 kHz to 110 GHz) and
power levels (–70 dBm to +20 dBm).
Table 13. Typical root sum of squares (rss) uncertainty on the calibration factor data printed on the power sensor
Frequency (GHz)
8483A
8481D
8485D
8487D
R8486D
Q8486D
0.0001
1.3
–
–
–
–
–
0.0003
1.2
–
–
–
–
–
0.001
1.1
–
–
–
–
–
0.003
1.2
–
–
–
–
–
0.01
1.2
–
–
–
–
–
0.03
1.2
–
–
–
–
–
0.05
1.2
–
–
–
–
–
0.1
1.2
–
–
–
–
–
0.3
1.2
–
–
–
–
–
1
1.2
0.8
1.4
1.3
–
–
2
1.2
0.8
1.4
1.3
–
–
4
–
0.8
1.7
1.4
–
–
6
–
0.9
1.7
1.4
–
–
8
–
1.0
1.7
1.4
–
–
10
–
1.1
1.9
1.5
–
–
12
–
1.2
1.9
1.5
–
–
14
–
1.1
2.0
1.6
–
–
16
–
1.5
2.1
1.7
–
–
18
–
1.7
2.2
1.7
–
–
22
–
–
2.7
1.9
–
–
26.5
–
–
2.8
2.2
3.0
–
28
–
–
2.9*
2.3
3.2
–
30
–
–
3.2*
2.4
3.0
–
33
–
–
3.3*
2.6
3.0
4.2
34.5
–
–
–
2.6
3.0
4.2
37
–
–
–
2.7
3.0
4.2
40
–
–
–
3.0
–
4.2
42
–
–
–
3.2
–
4.9
44
–
–
–
2.5
–
5.1
46
–
–
–
3.8
–
5.5
48
–
–
–
3.8
–
5.8
50
–
–
–
5.0
–
6.2
* These uncertainties only apply to Option 033.
** The 8480 Series sensors in the table do not include discontinued models.