E-series e9300 average power sensor specifications – Atec Agilent-EPM Series User Manual
Page 14
14
E-Series E9300 Average Power Sensor Specifications
The E-Series E9300 wide dynamic
range, average power sensors are
designed for use with the EPM family
of power meters. These specifications
are valid ONLY after proper calibration
of the power meter and apply for CW
signals unless otherwise stated.
Specifications apply over the tem-
perature range 0 °C to 55 °C unless
otherwise stated, and specifications
quoted over the temperature range
25 °C ±10 °C, conform to the standard
environmental test conditions as
defined in TIA/EIA/ IS-97-A and
TIA/EIA/IS-98-A.
Table 7. E9300 Series two-path specification
“A” suffix sensors
“B” suffix sensors
“H” suffix sensors
High power path
–10 to +20 dBm
+20 to +44 dBm
0 to +30 dBm
Low power path
–60 to –10 dBm
–30 to +20 dBm
–50 to 0 dBm
Table 8. E9300 Series sensors specification
Model
Frequency
range
Maximum SWR
(25 °C ± 10 °C)
Maximum SWR
(0 to 55 °C)
Maximum power
Connector
type
–60 dBm to +20 dBm wide dynamic range sensors
E9300A
10 MHz to 18 GHz
10 MHz to 30 MHz: 1.15
30 MHz to 2 GHz: 1.13
2 GHz to 14 GHz: 1.19
14 GHz to 16 GHz: 1.22
16 GHz to 18 GHz: 1.26
10 MHz to 30 MHz: 1.21
30 MHz to 2 GHz: 1.15
2 GHz to 14 GHz: 1.20
14 GHz to 16 GHz: 1.23
16 GHz to 18 GHz: 1.27
+25 dBm (320 mW) average;
+33 dBm peak (2 W)
(< 10 μsec)
Type-N (m)
E9301A
10 MHz to 6 GHz
10 MHz to 30 GHz: 1.15
30 MHz to 2 GHz: 1.13
2 GHz to 6 GHz: 1.19
10 MHz to 30 MHz: 1.21
30 MHz to 2 GHz: 1.15
2 GHz to 6 GHz: 1.20
+25 dBm (320 mW) average;
+33 dBm peak (2 W)
(< 10 μsec)
Type-N (m)
E9304A
9 kHz to 6 GHz
9 kHz to 2 GHz: 1.13
2 GHz to 6 GHz: 1.19
9 kHz to 2 GHz: 1.15
2 GHz to 6 GHz: 1.20
+25 dBm (320 mW) average;
+33 dBm peak (2 W)
(< 10 μsec)
Type-N (m)
–30 dBm to +44 dBm wide dynamic range sensors
E9300B
10 MHz to 18 GHz
10 MHz to 8 GHz: 1.12
8 GHz to 12.4 GHz: 1.17
12.4 GHz to 18 GHz: 1.24
10 MHz to 8 GHz: 1.14
8 GHz to 12.4 GHz: 1.18
12.4 GHz to 18 GHz: 1.25
0 to 35 °C: 30 W avg
35 to 55 °C: 25 W avg
< 6 GHz: 500 W pk
> 6 GHz: 125 W pk
500 W.μS per pulse
Type-N (m)
E9301B
10 MHz to 6 GHz
10 MHz to 6 GHz: 1.12
10 MHz to 6 GHz: 1.14
0 to 35 °C: 30 W avg
35 to 55 °C: 25 W avg
< 6 GHz: 500 W pk
> 6 GHz: 125 W pk
500 W.μS per pulse
Type-N (m)
–50 dBm to +30 dBm wide dynamic range sensors
E9300H
10 MHz to 18 GHz
10 MHz to 8 GHz: 1.15
8 GHz to 12.4 GHz: 1.25
12.4 GHz to 18 GHz: 1.28
10 MHz to 8 GHz: 1.17
8 GHz to 12.4 GHz: 1.26
12.4 GHz to 18 GHz: 1.29
3.16 W avg
100 W pk
100 W.μS per pulse
Type-N (m)
E9301H
10 MHz to 6 GHz
10 MHz to 6 GHz: 1.15
10 MHz to 6 GHz: 1.17
3.16 W avg
100 W pk
100 W.μS per pulse
Type-N (m)
The E-Series E9300 power sensors
have two independent measurement
paths (high and low power paths) as
shown in Table 7.