beautypg.com

2 photometry, Photometry – YSI pHotoFlex STD Colorimeter User Manual

Page 68

background image

Technical data

pHotoFlex

®

STD

68

ba76122e01

02/2013

7.2

Photometry

Applicable directives

and standards

EMC

EC directive 89/336/EEC

EN 61326-1/A3:2003

FCC Class A

Meter safety

EEC directive 73/23/EEC

EN 61010-1 :2001

UL3101-1

CAN/CSA-C22.2 No. 1010.1-92

Climatic class

VDI/VDE 3540

IP protection class

EN 60529:1991

Optical measuring prin-

ciple

LED photometer with filter

Interference filter

436 nm, 517 nm, 557 nm, 594 nm, 610 nm, 690 nm

Accuracy:

± 2 nm

Photometric reproduc-

ibility

0.005 or better

Photometric resolution

0.001

Warm-up time

none

Measuring time

approx. 2s

Measured parameters

Concentration (method dependent, selectable display type), absor-
bance, transmission

Measuring range

Absorbance:

-0.200 ... +2.000

Transmission:

1 ... 150 %

User-defined programs

100

Resolution Transmis-

sion

1.00 ... 9.99

0.01 %

10.0 ... 150

0.1 %