What to do if...

pHotoFlex

®

STD

66

ba76122e01

02/2013

6.2

Photometry

Measuring range

undercut or exceeded

Obviously incorrect

measured values

Cause

Remedy

– Program not suitable

– Select program with suitable

measuring range

– Dilute the sample

Cause

Remedy

– Measurement disturbed by

external light

– Close the external light cover.

– Cell not correctly inserted

– Insert the cell so that it is

positioned on the bottom of
the cell shaft.

– Cell contaminated

– Clean the cell

– Cell shaft contaminated

– Clean the cell shaft

– Dilution set incorrectly

– Set the dilution

– Selected program unsuitable

– Select other program

– Zero measurement incorrect

– Perform zero measurement

– Blank value incorrect

– Remeasure the blank value