Administrating user-defined methods
4.11.1 Options for method administration
User-defined programs can be
entered
read out
erased.
To store user-defined programs (methods) in the meter, determine a
calibration line for your method yourself and transmit the data of this
calibration line to the pHotoFlex
®
STD.
Administration of the method data can be carried out in the LSdata soft-
ware (see LSdata operating manual) or with the aid of a terminal pro-
gram ( see section 4.11.2).
You need the following data in any case:
Data
Possible entries/examples
Program No:
900 ... 949
Model:
Any name (max. 5 characters)
Cell [mm]:
16 or 28
Wave length [nm]:
436, 517, 557, 594, 610 or 690
MRB (measuring range start):
e.g. 0.05
MRE (measuring range end):
e.g. 8.00
E0 (reagent blank value):
e.g. 0.0
Slope: e.g.
1.0
Resolution:
0.0001, 0.001, 0.01, 0.1, 1, 2, 5,
10 or 100
Citation form:
e.g. PO4-P (max. 9 characters)
Unit:
e.g. mg/l (max. 7 characters)
Blank required:
(No/Yes)
0 or 1
Standard adjustment possible:
(No/Yes)
0 or 1
Standard adjustment required:
(No/Yes)
0 or 1