YSI pHotoFlex STD Colorimeter User Manual
Page 40
Operation
pHotoFlex
®
STD
40
ba76122e01
02/2013
If a standard adjustment is stored for a program, the standard adjust-
ment is automatically used for measurement. The usage of the stan-
dard adjustment is documented together with the measured value and
indicated in the measured value display with [Cal].
6
Enter the nominal value of the standard.
Note:
Enter the decimal separator with
7
Confirm the entered nominal value with
8
Insert a cell with standard (see section 4.2).
9
Start the measurement of the standard with
After measuring, the result of the standard adjustment is dis-
played and stored.
As the result, the measured value and the adjustment (in %) or
Error is displayed.
10
Confirm the result with
The standard adjustment is completed.
The meter is ready to measure.
or:
Discard the result with
Subsequently, carry out a new standard adjustment.
mg/l Cu
_
Enter nominal value
i304: Cu-1 TP
Cell
i = 10 mm
i 0.600 mg/l Cu:
2.000 (93.2%)
Accept
Measure standard
- 3100 (50 pages)
- 3200 (86 pages)
- 5000 (70 pages)
- 5200 (148 pages)
- 5200A (128 pages)
- 5200A (121 pages)
- 5400 (127 pages)
- 5239 Dissolved Oxygen Probe (23 pages)
- 55 (26 pages)
- 550A (18 pages)
- 556 MPS - Quick Start Guide (4 pages)
- 58 (40 pages)
- 5905 (22 pages)
- 6600V2 - User Manual (379 pages)
- 60 (33 pages)
- 63 (46 pages)
- 650 MDS (72 pages)
- 9300 Photometer (11 pages)
- 9500 Photometer (143 pages)
- 9500 Photometer (18 pages)
- CR 3200 Thermoreactor (42 pages)
- CR 4200 Thermoreactor (42 pages)
- CR2200 Thermoreactor (36 pages)
- DO200 (40 pages)
- EC300 (40 pages)
- EC300A (40 pages)
- EC30A (12 pages)
- IFL 700 IQ (46 pages)
- IQ SensorNet 2020 XT Terminal (196 pages)
- IQ SensorNet FDO Sensor (48 pages)
- IQ SensorNet MIQ/A-MOD Module (42 pages)
- IQ SensorNet MIQ/A-PR Module (36 pages)
- IQ SensorNet MIQ/IC2 Module (30 pages)
- IQ SensorNet MIQ/IF232 Module (18 pages)
- IQ SensorNet MIQ/JB Module (18 pages)
- IQ SensorNet MIQ/JBR Module (20 pages)
- IQ SensorNet MIQ/PS Module (22 pages)
- IQ SensorNet MIQ/R6 Module (56 pages)
- IQ SensorNet NitraLyt Sensor (64 pages)
- IQ SensorNet SensoLyt Sensor 700 IQ (SW) (46 pages)
- IQ SensorNet Sensors Encoded Fieldbus Data (12 pages)
- IQ SensorNet TetraCon Sensor 700 IQ (SW) (32 pages)
- IQ SensorNet ViSolid Sensor 700 IQ (60 pages)
- MultiLab 4010-1 (108 pages)
- MultiLab 4010-2 (138 pages)