8 programs, Programs, Section 4.5.8) – YSI pHotoFlex STD Colorimeter User Manual
Page 43
pHotoFlex
®
STD
Operation
43
ba76122e01
02/2013
If
iCalibration error! was displayed as the calibration result, a note au-
tomatically reminds you of another zero adjustment before the next
measurement.
Measuring is not possible without a valid zero adjustment.
4.5.8
Programs
Displaying program
data
You can view the most important data of all methods.
The method data is ordered according to the program number.
4
Insert the cell (see section 4.2).
5
Set another cell with <> <> and
essary.
6
Start the measurement of the zero adjustment with
After measuring, the result of the zero adjustment is displayed
and stored.
iZero adjust. successful! (successful zero adjustment) or
iCalibration error! (zero adjustment not successful) is dis-
played as the result.
The zero adjustment is completed.
7
Confirm the result with
The zero adjustment is completed.
The meter is ready to measure.
iInsert zero
cell (dist. water)
Cell
16 mm
Start measurement
Zero adjust.
1
Open the Configuration / Photometry / Programs menu.
The display shows the most important data of the selected pro-
gram.