beautypg.com

8 programs, Programs, Section 4.5.8) – YSI pHotoFlex STD Colorimeter User Manual

Page 43

background image

pHotoFlex

®

STD

Operation

43

ba76122e01

02/2013

If

iCalibration error! was displayed as the calibration result, a note au-

tomatically reminds you of another zero adjustment before the next
measurement.
Measuring is not possible without a valid zero adjustment.

4.5.8

Programs

Displaying program

data

You can view the most important data of all methods.
The method data is ordered according to the program number.

4

Insert the cell (see section 4.2).

5

Set another cell with <> <> and as nec-
essary.

6

Start the measurement of the zero adjustment with ENTER>.

After measuring, the result of the zero adjustment is displayed
and stored.
iZero adjust. successful! (successful zero adjustment) or

iCalibration error! (zero adjustment not successful) is dis-
played as the result.
The zero adjustment is completed.

7

Confirm the result with .

The zero adjustment is completed.
The meter is ready to measure.

iInsert zero
cell (dist. water)

Cell

16 mm

Start measurement

Zero adjust.

1

Open the Configuration / Photometry / Programs menu.

The display shows the most important data of the selected pro-
gram.