VEGA VEGAPULS 61 (≥ 2.0.0 - ≥ 4.0.0) Foundation Fieldbus Approval according to LPR radio standard User Manual
Page 69
-
Max. spectral transmission power den-
sity
+6
dBm/50 MHz EIRP
-
Max. power density in a distance of 1 m
< 1 µW/
cm²
Ambient conditions
Ambient, storage and transport temperature
-40 … +80 °C (-40 … +176 °F)
Process conditions
For the process conditions, please also note the specifications on the type label. The lower value
always applies.
Vessel pressure
-
Encapsulated antenna system
-1 … 3 bar (-100 … 300 kPa/ -14.5 … 43.5 psi)
-
Plastic horn antenna
-1 … 2 bar (-100 … 200 kPa/-14.5 … 29.0 psig)
-
Version with adapter flange from DN 150
PP
-1 … 1 bar (-100 … 100 kPa/-14.5 … 14.5 psig)
Process temperature (measured on the
process fitting)
-40 … +80 °C (-40 … +176 °F)
Vibration resistance
-
With adapter flange
2 g at 5 … 200 Hz according to EN60068-2-6
(vibration with resonance)
-
with mounting strap
1 g at 5 … 200 Hz according to EN60068-2-6
(vibration with resonance)
Shock resistance
100 g, 6 ms according to EN60068-2-27 (mechan-
ical shock)
Electromechanical data - version IP 66/IP 67 and IP 66/IP 68; 0.2 bar
Cable entry/plug
4)
-
Single chamber housing
l
1 x cable gland M20 x 1.5 (cable: ø 5 … 9 mm),
1 x blind stopper M20 x 1.5
or:
l
1 x closing cap M20 x 1.5; 1 x blind stopper
M
20 x 1.5
or:
l
1 x closing cap ½ NPT, 1 x blind plug ½ NPT
or:
l
1 x plug (depending on the version), 1 x blind
stopper M20 x 1.5
-
Double chamber housing
l
1 x cable entry M20 x 1.5 (cable: ø 5 … 9 mm),
1 x blind stopper M20 x 1.5; 1 x blind stopper
M
16 x 1.5 or optional
5)
1 x plug M12 x 1 for
external indicating and adjustment unit
or:
4)
Depending on the version M12 x 1, according to ISO 4400, Harting, 7/8" FF.
5)
Only with Ex d version
VEGAPULS
61 • Approval according to LPR radio standard
69
11 Supplement
41716
-EN
-120301