6 Set up with the display and adjustment module
MINITRAC 31 • Profibus PA
43836-EN-131119
This menu item appears only if you have selected under adjustment
mode (setup/adjustment mode) the "Two-point adjustment".
In this menu item you can set the max. pulse rate (ct/s) at which the
sensor should switch over.
Empty the vessel until the MINITRAC 31 is uncovered.
You thus get the max. pulse rate (ct/s) for the "uncovered" adjustment.
Enter the requested pulse rate manually or let the rate be determined
by MINITRAC 31. Automatic determination of the pulse rate should be
given preference.
You can enter the adjustment point (ct/s) manually.
You can let the adjustment point be determined by MINITRAC 31.
Since the adjustment is very comprehensive, the menu points of
Function Blocks 1 (FB1) were put together in a submenu.
In menu item"Channel" you determine which measured value the
output refers to.
In menu item "Scaling unit" you define the scaling variable and the
scaling unit for the pulse rate on the display, e.g. volume in percent.
Adjustment "uncovered"
(two-point adjustment)
Setup - AI FB1
Setup - AI FB1 - Channel
Setup - AI FB1 - Scaling
unit