Setup - Adjustment with the PC

• Click in the window "

Meas. environment“ to

"Measuring conditions“.

• In the menu window "

Measuring conditions“

you click on the options corresponding to
your application.

• Confirm with

"OK“.

After a few seconds of saving during which
the adjustments are permanently saved in
the sensor, you are again in the window
"

Meas. environment“.

Sonic velocity

When measuring in a gas composition devi-
ating from air, e.g. pure nitrogen, the sonic
velocity will change. Without correction, this
would distort the measuring result.

In the menu "

Sonic velocity“ the sonic velocity

can be adapted to the correct value in case
of another gas composition.

• Confirm with "

OK“.

• Click in the menu window "

Meas. environ-

ment“ to

"Quit“.

You are again in the menu window "

Sensor

optimisation“.

Echo curve

If, due to vessel installations, you expect
strong false echoes, a correction (if possible)
of the mounting location and orientation (dur-
ing simultaneous monitoring of the echo
curve) can help localise the false echoes and
reduce their intensity. With the menu item
"

Echo curve“ in the menu window "Sensor

optimisation“ you can see the form and the
intensity of the detected ultrasonic echoes.

Meas. environment/Meas. conditions