19
Technical data
Measuring range
(reference plane is the transducer end. On VEGASON 54 … 56 in version A the lower flange
side is the reference plane.)
VEGASON 51
- liquids
0.25 … 4 m
- solids
0.3 … 2 m
VEGASON 52
- liquids
0.4 … 7 m
- solids
0.5 … 3.5 m
VEGASON 53
- liquids
0.6 … 15 m
- solids
0.75 … 7 m
VEGASON 54 in general
1.0 … 25 m
VEGASON 55 in general
0.8 … 45 m
VEGASON 56
- version A
1.8 … 70 m
- version B … D
1.4 … 70 m
Output signal of the sensors
Signal output of the sensors
digital output signal in two-wire technology
(VBUS): the digital output signal (meas.
signal) is superimposed power supply of
VEGASCAN and further processed in
VEGASCAN
Meas. data inputs on VEGASCAN
Number of inputs
3 meas. data inputs on VEGASCAN 850B
6 meas. data inputs on VEGASCAN 850C
Supply voltage from VEGASCAN
approx. 24 V
Output current per meas. data input
max. 1 A
Output power
per meas. data input
max. 23 W (15 W continuous power)
Resistance of the signal cable
max. 15 W per wire or
max. 1000 m cable length
Integration time
0 … 999 seconds (selectable in the sensor)
0 … 600 seconds (selectable in VEGASCAN)
Two-wire technology: The digital output signal (meas. signal) and the power supply are led
through one cable.