beautypg.com

3 adjustment with the hart® handheld, 3 adjustment with the hart, Handheld – VEGA VEGADIF 34 … 51 User Manual

Page 38

background image

38

VEGADIF 34 … 51

20094-EN-030731

250

Ri < 250

+

-

Ri > 250

+

-

Electronics powered by PLC with active input circuit

Set-up

PLC

PLC

4.3 Adjustment with the HART

®

handheld

Apart from adjustment with the 4-key adjustment element in the sensor and adjustment with the
PC and the adjustment software VVO, a VEGADIF can be also operated with a HART

®

handheld.

If the inner resistance of the power supply is less than 250

, a resistor of

R

X

= 250

1

/

4

W must be connected to the signal /connection cable during adjustment.

The digital adjustment and communication signals would otherwise be short-circuited because
of insufficient resistances of the supply current source or the processing system - sensor com-
munication would not be ensured.