5 error codes – VEGA VEGAMET 614V User Manual
Page 46
46
VEGAMET 614V
Diagnosis
9.5 Error codes
Error code
Meaning
E004
This operation is not supported by the instrument (different software versions).
You have tried to activate a function with a newer software version of VEGA
Visual Operating in an instrument with older software version. However, the
instrument cannot support this operation.
Have the software of your instrument updated.
E007
Sensor type does not match.
A VBUS sensor with the serial number given by you was found, however the
sensor type does not match the settings.
Check the entered serial number and modify the configuration.
E008
VBUS sensor not found. Sensor cannot be found.
Probable reasons:
- no sensor connected
- no sensor with this serial number connected
- line break (VBUS)
- short-circuit (VBUS)
- sensor is defective
- interfered communication (VBUS)
Also possible:
- failure on the VBUS transmitter/receiver part of VEGAMET
Check the serial number, the connection cable, the sensor and if necessary
VEGAMET.
E010
The serial number of the connected VBUS sensors are not yet assigned to an
input component.
E013
VBUS sensor signals failure; however the communication with the sensor is
okay.
Probable reasons:
- failure on the VBUS sensor
Check the VBUS sensor.
E016
Empty/full adjustment reversed. The values for the empty adjustment are
above the values for the full adjustment. Carry out the adjustment again.
E017
Adjustment span too small. The adjustment values are too close together or are
identical. Carry out the adjustment again.
E021
Span of the scaling values too small or values reversed. The scaling values are
too close together or are identical. Please check the scaling.