Sensor optimization

40

71 - 1 … 75 - 1

Information for sensor optimization

If optimization is required, always compare the local
conditions with the programmed parameters - if
necessary correct or adapt.

Basic test of

Mode-no.

Information

Selection / measures

Page

1 - 09

product type

liquids or solids

p. 42

1 - 10

vessel geometry

vessel top flat or dished

p. 42

1 - 10

solid

define particle size /

p. 43

1 - 11

angle of repose

1 - 19

rate of level change

define in cm / sec.

p. 46

False echo tests

Mode-no.

Mode description

Page

False echo BEFORE useful echo (output too high)

1 - 07

start of working range

define

p. 41

1 - 14

17

false echo

learn (activate)

p. 45

1 - 18

profile interpolation

activate

p. 46

False echo AFTER useful echo (output too low)

1 - 08

end of working range

define

p. 41

1 - 12

/

13

multiple false echo

reduce

p. 44

1 - 21

/

22

max. gain

limited

p. 47

Supplementary programmings

1 - 23

measuring window

switch on or off

p. 48

1 - 24

false echo

determine

p. 48

1 - 25

real distance

program

p. 49

1 - 27

basic adjustment

define

p. 49