Supplementary programmings
71 - 1 … 75 - 1
37
Basic adjustment
MODE 0 - 97
After having finished the parameter adjustment of
one or several modes it may be necessary to reset
the parameters of these modes to factory preset
values.
Mode 0 - 97 offers this possibility. Previously
adjusted parameters are cancelled. Cancellation is
indicated with CAL on the display (approx. 5 secs.).
Programming example ..........................................
• select mode 0 - 97 in MODEFIELD
• program figure 1 in PARAMETER FIELD
• then store with STO-button, CAL is displayed for
5 secs. in the PARAMETER FIELD.
Keyword
MODE 0 - 98
Description
Acc. to the factory preset the keyword is switched
off and therefore data input is adjustable.
All mentioned modes can be selected as described
before and their parameters indicated on the
display of the sensors and perhaps modified.
The activation of the keyword (mode 0 - 98 = 1)
ensures the protection of the parameters for unau-
thorized modifications.
Adjustment of the data input is possible only after
programming the keyword (over writing of the indi-
cated key symbol).
This programming can be carried out in each mode,
the whole mode range is then accessible (mode 98
will be 0).
The keyword is:
0070
Activate keyword ......................................................
• select mode 0 - 98 in MODEFIELD
• program figure 1 in PARAMETER FIELD
• then store with STO-button
Adjust keyword .........................................................
• select the mode requested for parameter adjust-
ment, here in the example mode 0 - 43 in
MODEFIELD
• select data input, the key symbol
0 – –
n is indi-
cated in the PARAMETER FIELD
• program
0070
as keyword
• then store with STO-button
• the parameter adjustment of e.g. mode 0 - 43 can
be continued