Measurement, Current output

34

71 - 1 … 75 - 1

MODE 0 - 65 and 0 - 66

Output characteristic

for current output 1 and 2

The pulse-echo measuring system can be equipped
with a current module. The module has two current
outputs.

The characteristic of the current output is fixed by
an initial and final point.

The factory preset defines a course 4 … 20 mA
acc. to 0 … 100 %.

Within the total range of 0.00 … 20.00 mA each
current value can be programmed for a raising or
falling level.

The current-

between initial and final value must

be min. 1 mA.

• select mode 0 - 65 in MODEFIELD
• program 16.00 mA in PARAMETER FIELD
• then store with STO-button

• select mode 0 - 66 in MODEFIELD
• program 00.00 mA in PARAMETER FIELD
• then store with STO-button

%
100

0

mA
20

4

End of characteristic
100 % ^ 16 mA

Start of characteristic
0 % ^ 0 mA

Factory preset

Programming example